LC-MS/MS limits of quantification (LOQ).

Serum #72 Q Low Serum #72 Q High Serum #73 Q Low Serum #73 Q High LOQ Range
Mean (nM) RSD (%) Recovery (%) Mean (nM) RSD (%) Recovery (%) Mean (nM) RSD (%) Recovery (%) Mean (nM) RSD (%) Recovery (%) (nm) (nM)
Estrone 1-sulfate 0.43 7 105 0.94 5.7 99 0.39 8.7 96 0.92 8.9 96 0.026 LOQ-10
Cortisone 4.63 0.87 100 10.6 7.6 98 4.2 4.1 91 10.4 3.4 97 0.19 LOQ-112
Cortisol 35.3 8.1 102 81 2.3 100 32.3 0.9 93 77.7 1.7 96 1.9 LOQ-794
Dehydroepiandrosterone sulfate 720 7.5 104 1652 2.7 102 671 1.1 97 1584 1.4 98 19 LOQ-3000
Corticosterone 38 7.3 107 46.5 12 109 18.3 8.7 103 25.5 0.54 98 0.1 LOQ-144
11-Deoxycotisol 1.86 2.3 104 4.07 6.8 98 1.78 1.9 96 4.02 3.4 95 0.017 LOQ-40
Δ4-androstenedione 2.13 0.74 102 4.72 4.2 97 3.39 6.8 94 5.98 3 93 0.042 LOQ-1746
Testosterone 5.62 1.1 111 7.71 3.1 99 49.3 3 106 51.4 3 98 0.012 LOQ-1732
17α-hydroxyprogesterone 2.09 11 96 4.53 7 95 2.18 6.3 90 4.67 0.36 93 0.1 LOQ-1513
Progesterone 3.48 1.3 100 7.99 4.9 98 3.89 5.5 92 8.33 5.2 94 0.036 LOQ-500

LC-MS/MS limits of quantification (LOQ) and range of calibration curves were based on 10 standards prepared as for human serum analysis (20). (B) LC-MS/MS validation: Inter day control materials for this study were prepared in two pools of mouse serum (serum #72 and #73) spiked in low (Q Low) and high (Q High) levels. Results are the mean (n=3) of control materials analyzed in three batches.