Method error and reliability for the repeated measurements in acoustic rhinometry tested in an identical way in standing position and in sitting position. Mean difference: the second measurement subtracted from the first measurement

Mean
difference
Standard
deviation
Method error
(s(i))
Reliability
(Houston)
Standing position, right nostril

Calculated resistance (cm H2O/L/min) -0.477 6.969 0.055 1

Volume (cm3) -0.013 3.033 0.001 1

Minimum cross-sectional area (MCA, cm2) 0.005 0.177 0.001 1

Distance to MCA (cm) -0.058 1.445 0.007 1

Standing position, left nostril

Calculated resistance (cm H2O/L/min) -1.328 4.791 0.152 0.999

Volume (cm3) 0.173 1.539 0.02 1

Minimum cross-sectional area (MCA, cm2) 0.027 0.116 0.003 0.999

Distance to MCA (cm) -0.114 0.552 0.013 0.999

Sitting position, right nostril

Calculated resistance (cm H2O/L/min) -0.396 3.593 0.048 1

Volume (cm3) 0.594 1.934 0.072 0.999

Minimum cross-sectional area (MCA, cm2) 0.035 0.135 0.004 0.999

Distance to MCA (cm) -0.156 0.936 0.019 1

Sitting position, left nostril

Calculated resistance (cm H2O/L/min) -3.053 15.616 0.37 0.999

Volume (cm3) 0.57 1.837 0.069 0.999

Minimum cross-sectional area (MCA, cm2) 0.043 0.149 0.005 0.999

Distance to MCA (cm) -0.29 0.793 0.035 0.999

MCA = minimum cross-sectional areas.