Skip to main content
PLOS One logoLink to PLOS One
. 2021 Feb 11;16(2):e0245334. doi: 10.1371/journal.pone.0245334

Assessment of breast cancer surgical margins with multimodal optical microscopy: A feasibility clinical study

Mark T Scimone 1, Savitri Krishnamurthy 2, Gopi Maguluri 1, Dorin Preda 1, Jesung Park 1, John Grimble 1, Min Song 1, Kechen Ban 1, Nicusor Iftimia 1,*
Editor: Matthew Bogyo3
PMCID: PMC7877783  PMID: 33571221

Abstract

Providing surgical margin information during breast cancer surgery is crucial for the success of the procedure. The margin is defined as the distance from the tumor to the cut surface of the resection specimen. The consensus among surgeons and radiation oncologists is that there should be no tumor left within 1 to maximum 2 mm from the surface of the surgical specimen. If a positive margin remains, there is substantial risk for tumor recurrence, which may also result in potentially reduced cosmesis and eventual need for mastectomy. In this paper we report a novel multimodal optical imaging instrument based on combined high-resolution confocal microscopy-optical coherence tomography imaging for assessing the presence of potential positive margins on surgical specimens. Since rapid specimen analysis is critical during surgery, this instrument also includes a fluorescence imaging channel to enable rapid identification of the areas of the specimen that have potential positive margins. This is possible by specimen incubation with a cancer specific agent prior to imaging. In this study we used a quenched contrast agent, which is activated by cancer specific enzymes, such as urokinase plasminogen activators (uPA). Using this agent or a similar one, one may limit the use of high-resolution optical imaging to only fluorescence-highlighted areas for visualizing tissue morphology at the sub-cellular scale and confirming or ruling out cancer presence. Preliminary evaluation of this technology was performed on 20 surgical specimens and testing of the optical imaging findings was performed against histopathology. The combination of the three imaging modes allowed for high correlation between optical image analysis and histological ground-truth. The initial results are encouraging, showing instrument capability to assess margins on clinical specimens with a positive predictive value of 1.0 and a negative predictive value of 0.83.

1. Introduction

In breast cancer cases that are candidates for breast-conserving surgery, it is extremely important to achieve negative margins around the primary tumor. The margin is defined as the distance from the tumor to the cut surface of the resection specimen. The consensus among surgeons and radiation oncologists is that there should be no tumor left within 1 to maximum 2 mm from the surface of the surgical specimen [13] in patients with ductal carcinoma in situ and no tumor at the inked margin in patients with invasive mammary carcinoma. If a positive margin remains, there is substantial risk for tumor recurrence [4, 5]. Published reports indicate a 20–70% rate of positive margins left after lumpectomy [68]. The required repeat surgery in these cases may result in potentially reduced cosmesis and eventual need for mastectomy if cancer recurrence is not rapidly detected.

Current techniques for intraoperative pathologic assessment involve touch prep and frozen section analysis [9, 10]. Touch-prep or imprint cytology allows for cytologic evaluation of the whole lumpectomy surface and has good sensitivity and specificity [11]. However, this technique is time consuming, is cumbersome, requires special expertise, and does not detect tumor foci close to the lumpectomy surface (residual cells < 2mm from the margin). Frozen section analysis is a technically challenging procedure due to the significant amount of fatty tissue found in breast specimens, which makes it difficult to freeze the tissue effectively. While the best results using this technique have been shown to reduce the rate of second operations to about 20% [12], false negatives still occur at high frequency [13]. Due to these difficulties, the vast majority of hospitals in the U.S. do not perform intra-operative margin assessment.

Clinical research approaches for intraoperative margin assessment of breast specimens include micro-CT, MRI, high frequency ultrasound, positron emission tomography (PET), and radiofrequency spectrometry. MRI and micro CT do not provide a cellular-level picture of the tissue microstructure that is necessary to assess margins accurately and therefore their sensitivity is not very high [1416]. Ultrasound guidance for surgical margins can be useful for determining margin status in a low-cost manner, especially for women with dense breasts where other methods may not work as well. However, its reported sensitivity is rather modest (~44% in some studies) [17]. PET has shown a sensitivity of 92% to 96% and a specificity of 84% to 91% for breast margins assessment [18, 19]. However, PET is expensive, time-consuming, and invasive, which makes it not appealing for intraoperative margin assessment of breast surgical specimens. Radio-frequency spectrometry, has had limited sensitivity (60% or less) for positive margins and can only investigate relatively small surface areas (~2 mm) [20]. Limited sensitivity leads to margins left behind and thus to cancer recurrence.

Optical imaging techniques have been investigated for evaluation of physiological, chemical, and morphological changes associated with cancer, and so for facilitating margin assessment. These techniques include fluorescence imaging, Raman spectroscopy, optical coherence tomography (OCT), and photoacoustic tomography, among others [2128]. Although most of these techniques achieved high accuracy in cancer delineation, they have not yet been translated into routine clinical practice due to some inherent pitfalls, such as reduced sensitivity, limited speed, inability to quickly cover a large tissue area, and the requirement of an expert operator [29].

However, it has been noted that some of the optical modalities provide complementary information, and thus their combined use might help to improve sensitivity, and even speed for evaluation of large surface areas of surgical resection specimens. Therefore, this paper aims to introduce a novel multimodal microscopy device that might enable more accurate detection of breast tumor margins at a reasonable speed. By combining fluorescent microscopy (FM), optical coherence tomography (OCT), and reflectance confocal microscopy (RCM), the platform provides rapid delineation and high-resolution assessment of tumor margins with minimal feedback from the operator. Briefly, using a fluorescent cancer targeting agent, FM highlights regions-of-interest (ROIs) of likely positive margins. While providing a large field of view and relatively rapid analysis of the specimens, the specificity of FM is not exceedingly high (usually prone to false positives). Therefore, combined OCT/RCM imaging is used as well to analyze the ROI at submicron scale resolution and confirm or rule out cancer presence. RCM is used to determine if the fluorescence imaging highlighted margins are positive, while OCT is used to rapidly determine their extent in the lateral and deep aspects. In this way, the system is optimized for rapid acquisition time, keeping intraoperative utility in mind as a primary objective.

The multimodal instrument and software for image analysis developed have been evaluated on several breast tissues obtained from surgical resection specimens encountered in clinical practice. The detected margins were compared with the ground truth of histopathological examination of the imaged tissue. Twenty specimens were analyzed for this initial report. A strong correlation with histology (100%) was found in all specimens that presented positive cancer margins (8 of the total 20). The imaging of the benign specimens also correlated well with the histology. However, RCM-guided OCT was prone to false positives in a small number of specimens (two), particularly in high scattering areas of normal breast lobules.

2. Materials and methods

2.1 Fluorescence imaging agent

A contrast agent capable of highlighting areas of the specimen that have potential positive margins can be used with the proposed microscope. The main requirement is to have close to 100% sensitivity, as otherwise positive margins may be left behind. For this study, a novel contrast agent was developed by our team (see structure in Fig 1). It consists of an optically silent peptide substrate containing two (near-infrared; NIR) fluorochromes (Cy5.5), internally quenched. This agent was synthesized to track positive margins on breast lumpectomy specimens via specific urokinase plasminogen activator (uPA) cleavage.

Fig 1. Peptide cleavage and fluorescence release.

Fig 1

Thus, the multifunctional peptide can be cleaved by highly expressed breast cancer enzymes, like uPA, to provide a fluorescence signal and indicate potential cancer presence at the specimen surface. Once the quenched NIR agent is cleaved, the released mono-NIR dye species become fluorescent when activated by near-infrared light. To further increase specific targeting of cancer tissue, a peptide moiety targeting Her2 or other cancer receptors may be appended, as suggested in Fig 1. A spacer (aminohexanoic acid) may be used to ensure that this peptide moiety will maintain fast uPA cleavage kinetics. However, enhancement of cancer specificity was not a major focus of this preliminary study.

The targeting agent synthesized by our team showed 90% quenching efficiency and 10X fluorescence increase after cleavage (see Fig 2). The activation time was tested as well using variable concentrations of the uPA enzyme. Based on literature data, we estimated the concentration of the enzyme that is present in malignant tissue versus benign tissue and used these numbers as a baseline for our measurements. Literature reports [30] show that the average uPA level in benign tissue is 1 ng/mg protein, while in tumor tissue is 8 ng/mg protein. Considering an average volume of tissue of 5–10 cc, 1 mm penetration depth of the agent after incubation, 1 g/cc density (for breast tissue), and 1uM concentration of the contrast agent in saline solution, we determined that a ratio of 1:200 between the enzyme and substrate (E:S) is representative of benign tissue, while a ratio of 1:25 is representative of tumor tissue.

Fig 2. Contrast agent structure and fluorescence release full protease cleavage.

Fig 2

However, since the tumor tissue often infiltrates benign tissue, we also considered an intermediate ratio of 1:50. Using these ratios, we performed titration kinetic activation experiments to assess the fluorescence release as a function of time and E:S ratio. The results are outlined in Fig 3. As observed, the best contrast (4.5:1 for E:S = 25 vs. E:S = 200 and 3.5: for E:S = 50 vs. E:S = 200) was obtained after only 1 minute of incubation. After that, the cleavage begins at the lower (benign) concentration of uPA, resulting in approximately half the fluorescence increase as the malignant concentration. This suggests that that the optimum incubation time will be about 1 minute. However, to enable contrast agent diffusion in the tissue, our experiments showed that a maximum of 2 minutes incubation time is a good compromise between fluorescence contrast and agent diffusivity to about 1 mm deep. For longer incubation times, the smaller concentrations of the uPA enzyme present on the benign tissue start cleaving the agent, and thus degrade tumor contrast. After incubation, the unbound agent derived species are washed with saline to minimize any interference with the readout.

Fig 3. Contrast agent activation kinetics.

Fig 3

2.2 Multimodal imaging system

A simplified schematic of the optical setup is shown in Fig 4 and photographs of the instrument are shown in Fig 5.

Fig 4. Simplified schematic of the multimodal optical imaging microscope.

Fig 4

Fig 5. Photograph of the multimodal microscope.

Fig 5

A. General view; B- Inside view.

The FM subsystem consists of a high power LED Chip light source (Model Chanzon, 380nm-840nm / 3000mA), a highly sensitive NIR camera (Model PCO Edge 4.2, PCO Tech. Inc.), dichroic mirrors and band pass filters (Chroma Technologies) to properly match the excitation/emission bands of the CY 5.5 dye. The bandpass filter in front of the camera is positioned on a motorized mount, such that a brightfield image can also be taken, and overlaid by the NIR image. The field of view (FOV) of the fluorescence channel is 1.5 x 1.5 inches, which is sufficient to image many of the lumpectomy specimens. Following fluorescence ROI selection, the sample is automatically translated to be centered in the OCT/RCM objective field of view.

The RCM/OCT subsystem shares the same distal end optics (telescope lens and imaging objective), while in the upper stream the optical paths are separated by a dichroic mirror. Folding optics and telescope relay lenses are used in both the RCM and OCT optical paths such that the beam scanners can be relayed to the pupil plane of the imaging objective. The RCM subsystem is based on a typical confocal scheme. A resonant scanner (Model CR08, Cambridge Technologies, MA) and a galvanometer scanner (Model 6200H, Cambridge Technologies, MA) are used to generate a raster scan on the sample surface.

The two scanners are relayed to the imaging objective entrance pupil with two sets of relay lenses (telescope arrangement). The light from an 830 nm laser diode is directed to the scanners by a beam-splitter cube. The light returned from the sample is directed by the same beam-splitter cube to the avalanche photodiode (APD) detector. A pinhole is placed in front of the APD to reject the out-off-focus photons. Two polarizers are crossed in the emission and detecting paths to reduce the impact of sample specular reflections.

The OCT subsystem consists of a fiber optic (FO) interferometer with an optical delay line in the reference arm and a scanning engine in the sample arm. A superluminescent diode with 1310 nm central wavelength and approximately 100 nm 3dB bandwidth is used as the OCT light source. This provides a theoretical axial resolution of about 8 μm (in air), while the achieved resolution was 9 um due to imperfect balancing of the dispersion between the two arms of the interferometer. A fiber optic circulator is used to maximize light collection from the sample arm. Two galvanometer scanners (Model 6200H, Cambridge Technologies, MA), relayed to the pupil plane of the imaging objective, are used to generate a raster scan of the OCT beam on the sample surface.

Both the OCT and the RCM signals are digitized and processed by a System Control and Data Processing unit. A specially designed Vivascope frame grabber (Caliber ID, Rochester, NY) is used to digitize and process the RCM signals, enabling a display rate of 9 frames/sec. A graphical processing unit (GPU), model GTX 1060 with 1152 CUDA cores is used to expedite the OCT data processing (FFT, dispersion, and interpolation), allowing for 40 frames/sec real-time display.

2.3 Software for instrument control, data acquisition, and processing

The instrument control software was written in LabVIEW, employing C-based dynamic link libraries for rapid processing and easy debugging. A user-friendly graphical interface allows for instrument parameters setting, data acquisition and processing. The main steps for data collection and processing are illustrated in Fig 6.

Fig 6. Main steps for data acquisition and processing.

Fig 6

A: Bright field image of the specimen; B: False-color fluorescence imaging taken after specimen incubation with the contrast agent; C and C’: RCM images and zoomed in region shown in red box, respectively., D: OCT image of one B-scan; E automated segmentation of the OCT data through full volumetric image (one cross section shown here); F- Overlay of the positive margins on the specimen picture.

The bright field and subsequent fluorescence images of the specimen are first collected (see Fig 6A and 6B). The user defines a region of interest (ROI) that incorporates the fluorescence image, as shown in Fig 6B, and the instrument automatically scans this area to collect spatially co-registered cross-sectional OCT images and enface RCM images (see Fig 6D and 6C–6C’). Since the size of the ROI is usually larger than the instantaneous field of view for OCT and RCM (2.2 mm for OCT and 0.75 mm for RCM), a line-strip imaging approach is used. The number of strips and their length is automatically calculated by the software as a function of the ROI size. An overlap between adjacent strips is user-defined to enable the cross-correlation of the adjacent strips for full area image stitching.

Note that the RCM imaging over the entire area highlighted by FL imaging is not performed, as it would take too long on large areas. Only small regions of interest (ROIs) are used to determine if areas indicated by fluorescence imaging are cancer positive or not. Once these areas are confirmed, OCT imaging and segmentation is performed to determine the extent of the positive margins. Tissue-type classification is based on the analysis of the RCM image. With minimal user feedback (assignment of distinctive tissue-types on 2–3 locations of the RCM image), an automated algorithm (see Section 2.2) analyzes the collected images, determines tumor lateral and depth spreading on the OCT image (see cross-sectional tissue-type assignment on Fig 6E), and overlays the tumor margin on the specimen surface image (see Fig 6F).

2.4 Semi-automated software algorithm for tissue-type assignment

Following the collection and analysis of 3–5 RCM ROIs for each tissue type in different locations of the fluorescence highlighted area, such that both cancer and benign tissues are represented as determined by the pathologist, the overlapping enface OCT images are automatically assigned to the cancer and benign tissue types and used as training sets for the OCT image analysis algorithm. Representative training images are shown in Fig 7A.

Fig 7. Simplified representation of the automated algorithm for tissue differentiation.

Fig 7

A. Representative en-face and cross-sectional (b-scan) training images for adipose tissue and tumor tissue (benign representative images not shown for brevity). B. Principal component analysis of 9 spatial texture features shows some clustering in the 2 strongest components, but also considerable overlap. C. Singular value decomposition cumulative stair function showing that 5 or fewer parameters should enable strong classification capability.

Initial singular value decomposition and principal component analysis indicated that tissue slope and standard deviation of roughness are the most discriminating features.

The standard deviation of roughness and tissue slope were the most discriminating features, and their calculation is detailed further. Remarkably high frequency information was smoothed using a Gaussian kernel. Then, to calculate the roughness, the waviness profile was first determined and subtracted from the filtered image. Finally, a standard deviation filter was convolved with the roughness image. Tissue type assignment based off standard deviation of roughness was determined using values from the training image sets. The tissue slope was computed through the Prewitt image gradient method. Again, tissue-type assignment based on tissue slope was determined using values from the training image sets. All analysis was done in MATLAB, where the analysis and plotting take approximately 1 to 2 minutes, depending on the image size. The processing duration can be further shortened by implementing this code in C++.

Three-dimensional tissue assignment (in cross-sectional images, and in different z-planes of en-face), required depth adjusted image processing. That is, signal drop in depth was calculated for each image so that tissue-type assignment thresholds could account for it. Additionally, in cross-sectional images, the top and bottom surfaces needed to be segmented. The top surface is highly reflective and thus can be segmented by a peak-finding algorithm. The bottom (or maximum signal depth) was segmented by a histogram-based region-property analysis.

2.5 Tissue specimen collection and preparation

Breast surgical specimens were provided by MD Anderson Cancer Center. Twenty de-identified specimens were acquired from breast cancer patients undergoing surgical procedures (both lumpectomy and mastectomy) under the IRB approved protocol PA14-1036. The de-identified specimens utilized in the study were collected from breast surgical resections between 7/11/2019 and 11/13/2019 with waiver of informed consent. The MD Anderson Cancer Center Institutional Review Board approved the procurement of tissue for the study. Freshly collected specimens were incubated with the contrast agent for a maximum of 2 minutes, washed with saline twice, and placed on the microscope tissue holder. The specimens were slightly pressed against the microscope window facing the imaging objective, to maximize the area of the specimen from which images were collected. This custom tissue holder is particularly important for RCM, as the depth penetration requires the specimen surface to be relatively flat and close to the objective.

3. Results

The summary of the imaging and histopathology findings is presented in Table 1. Of the 20 specimens analyzed, histology showed positive cancer margins on 8 specimens and negative margins in the remaining 12. Fluorescence imaging showed positive margins in 10 of the 20 specimens, so all specimens with positive margins were identified. This is particularly important as no positive margins were missed. Fluorescence imaging was used to further guide RCM-OCT imaging. RCM imaging was used as input to the OCT segmentation and has correctly identified most of the tissue types, as it provided the resolution close to histopathology. Some errors were encountered in areas of normal lobules, which showed similar appearance to the cancer areas. OCT imaging correctly identified all 8 specimens with positive margins and showed false positive margins on other two specimens (mainly on areas of normal lobule locations, which showed similar scattering and image patterns like the cancer areas).

Table 1. Summary of imaging and histopathology findings.

20 Specimen study Fluorescence imaging RCM guided OCT segmentation Histopathology
Positive margins specimens 12 10 (2FP) 8 TP
Negative margins specimens 8 10 (0FN) 12 TN
RCM-guided OCT predictive values: PPV = 1.0; NPV = 0.83

Five representative examples of margin assessment on breast surgical specimens are shown in Figs 811. The five cases were chosen to show distinct types of specimens that might be found during surgery: 1) specimens with positive margins all around, 2) heterogeneous specimens with infiltrating tumor margins, 3) negative margin specimens that visually contain only adipose tissue at its surface, 4) heterogeneous benign specimens that after visual and tactile examination show areas similar to cancer, and 5) specimens with small deeper cancer foci, which may been missed by histopathology if the cuts are very close to specimen surface.

Fig 8. Multimodal optical imaging findings on a small surgical resection specimen (~14 mm x 8mm).

Fig 8

A. Camera image of the specimen on the microscope stage. B. Fluorescence imaging, where the green box shows the selected imaging ROI in the software, and the yellow-dotted line shows the area where there is higher fluorescence from the contrast agent. C. The histology slide, where red indicates invasive ductal carcinoma. D. Segmented enface OCT image showing cancer margins with corresponding X and Y cross-sectional images. Small red and yellow boxes show region of RCM images in panels E&F. Colored overlays shows algorithmic segmentation of tissues: tumor (red), adipose (green), and benign (blue). The cross-sectional images have segmented tumor overlaid in red. G. Projection of the cancer areas from the multiple OCT slices over a depth of approximately 1 mm. H. Overlay of the positive margins on the surgical specimen.

Fig 11. Example of negative margin identification and false positives on a heterogeneous benign specimen.

Fig 11

A) Camera photo of the specimen on the microscope stage. B) Fluorescence imaging seems to indicate an area of tumor in the top-right corner (yellow). C) Annotated histology indicating benign breast lobules (blue). D) Enface OCT image with corresponding X and Y cross-sectional images. Dotted-yellow line shows the plane of the cross-sectional images. Red and Yellow arrowheads show the region of RCM images in E and F panels. Colored overlays show algorithmic segmentation of tissues: tumor (red), adipose (green), and benign (blue). Colored boxes show region of RCM images in panels E & F. E) RCM image of lower part of the normal duct area. F) RCM image of adipose tissue. z-depth of RCM image is consistent at 100 μm. G) Enface OCT image with only the false positive algorithmic tumor predictions shown.

The first case (see Fig 8) is that of a fragment of a mastectomy specimen, which presented nearly complete positive margins, as it was not collected with the purpose of preserving positive margins. Imaging was performed on a single surface, since visually the specimen showed similar tissue nature all around. The goal of imaging this specimen was to determine the optical imaging/algorithm capability to correctly estimate cancer presence on the entire specimen surface, as indicated by histology. Indeed, as observed in the fluorescence image (Fig 8B), nearly the entire specimen surface showed enhanced fluorescent signal relative to the background, suggesting uniform cancer presence at the specimen’s surface. Cancer presence on over 80% of the specimen surface was confirmed by histology (see the red area annotated by the breast pathologist in Fig 8C). The enface segmented OCT image taken near specimen surface (see Fig 8D), as well the subsurface cancer presence, over a depth of approximately 1 mm, as shown projected on Fig 8G, confirm a similar cancer spreading area.

The second case, shown in Fig 9, is that of a small surgical specimen with positive margins (cancer infiltrating at the specimen’s surface). A fluorescence signal was noticed on one of the faces of the tissue specimen, as shown in Fig 9A. However, as confirmed by histology, the fluorescence area overestimated cancer presence, as the signal appears to be heightened in areas of adipose as indicated by the histopathology’s-annotated areas in the histology slide (see Fig 9C). The segmented enface OCT image taken near specimen surface (see Fig 9D), shows reasonable overlap with histology relative to cancer presence. Hhistology preparation led to loss of fatty areas towards the top of the specimen as the processing of the fatty tissue for histopathological sectioning is usually difficult. This is clear when comparing the histology to the OCT (Fig 9D and 9G) and camera image (Fig 9A). The cross-sectional segmentation and Fig 9G show that the area of cancer spreads as depth increases. This highlights the need for subsurface OCT imaging and volumetric segmentation. The RCM zoomed-in panels (Fig 9E and 9F) show the structure in a tumor and fat area, respectively.

Fig 9. Example of positive margins identification on a small lumpectomy specimen.

Fig 9

A. Camera image of the specimen on the microscope stage. B. Fluorescence imaging, where the red box shows the selected imaging ROI in the software, and the yellow-dotted line shows the area where there is higher fluorescence from the contrast agent. C. The histology slide, where red indicates invasive ductal carcinoma. D. Segmented enface OCT image with corresponding X and Y cross-sectional images. Dotted-yellow line shows the plane of the cross-sectional images. Colored overlays show algorithmic segmentation of tissues: tumor (red), adipose (green), and benign (blue), Colored boxes show region of RCM images in panels E-F. The z-depth shown in F corresponds to the imaging depth for all RCM panels G) Projection of the cancer areas from the multiple OCT slices over a depth of approximately 1 mm shows larger predicted tumor areas at depth than in panel D. H. Overlay of the positive margins on the surgical specimen.

The third example is that of a surgical specimen free of positive margins (see Fig 10), as indicated by the histology image (Fig 10C), which is shown to have only a benign breast lobule (yellow) and normal duct (green). The fluorescence imaging, however, indicated a small area of potentially positive margins (see Fig 10B, red asterisk). Confocal imaging of the fluorescence-highlighted area has indicated the presence of a benign lobule. Although in real clinical practice additional OCT imaging would be unnecessary when RCM indicates the absence of the positive margin, in this case we performed OCT imaging to confirm that the segmentation algorithm assigns this area to the correct tissue class. The enface OCT image (Fig 10D) of the specimen surface clearly shows the presence of fat cells on over 90% of the specimen surface, as well as of a small area of a benign breast lobule (see blue area on the right upper area of Fig 10D), as confirmed by RCM imaging (Fig 10E). Fig 10F confirms the presence of the fatty tissue on most of the specimen’s surface, as indicated by the segmented OCT image (Fig 10D).

Fig 10. Example of negative margin identification on a small lumpectomy specimen.

Fig 10

A) Camera photo of the specimen on the microscope stage. B) Fluorescence imaging seems to indicate positive margin presence (red asterisk). Green square indicates the imaging ROI selected in the software. C) Annotated histology indicating a benign breast lobule (yellow) and normal duct (green). D) Enface OCT image with corresponding X and Y cross-sectional images. Dotted-yellow line shows the plane of the cross-sectional images. Red and Yellow boxes show region of RCM images in E and F panels. Colored overlays show algorithmic segmentation of tissues: tumor (red), adipose (green), and benign (blue). E) RCM image of lower part of the normal duct area. F) RCM image of adipose tissue. z-depth of RCM image is consistent at 100 μm.

The fourth case is that of a heterogeneous specimen, where multiple tissue types are apparent by eye (Fig 11A). After incubation with the fluorescent contrast agent, a small region in the corner showed increased signal (Fig 11B), and thus warranted in depth analysis. The histological annotation shows the presence of multiple benign breast lobules (Fig 11C, blue circles). Again, these structural features contribute to small false positive tumor identification in the algorithmic segmentation. This is shown in Fig 11D and 11G. However, the RCM panel in Fig 11E was interpreted as benign by the histologist, showing how multiple imaging modes can inform a more accurate classification of the tissue.

The fifth case is that of a heterogeneous specimen, where multiple tissue types are apparent by eye (Fig 12A). After incubation with the fluorescent contrast agent, several regions showed increased signal (Fig 12B), and thus warranted in depth analysis. The histological annotation shows the presence cancer areas on the left and bottom areas of the specimen (red-annotated areas) (Fig 12C). However, this histology slide seems to be been taken close to the specimen surface, as cancer foci in other areas of the specimen were identified by optical imaging. The RCM panels in Fig 12F and 12G, indicate areas of admixed fibrous-tumor tissue (F) and fibrotic tissue (G). OCT image segmentation over the 1 mm depth has correctly indicated positive margins in areas indicated by histopathology, but also small cancer foci in areas indicated as fibro-adipose by histopathology. These areas might be real, as the histopathology was unfortunately performed at one single depth, closer to surface, while additional cancer foci might been present at higher depths.

Fig 12. Example of positive margin identification and potentially false negatives indicated by histopathology on a heterogeneous specimen.

Fig 12

A) Camera photo of the specimen on the microscope stage. B) Fluorescence imaging seems to indicate an area of tumor in the top-right corner (yellow). C) Annotated histology indicating cancer areas (red annotations). D) Enface segmented OCT image with corresponding X and Y cross-sectional images. Dotted-yellow line shows the plane of the cross-sectional images. E) En face OCT image. Red and Blue boxes on the OCT image show the regions of RCM images in the F and G panels. H) Overlay of the positive margins on the surgical specimen.

4. Discussion

A multimodal instrument was developed with the goal of examining breast surgical resections encountered in routine clinical practice for positive margin presence. This novel instrument combines three optical imaging modalities: fluorescence, confocal, and optical coherence tomography imaging. To enable enhanced contrast fluorescence imaging, a custom contrast agent was developed by attaching Cy5.5 to a quenched uPa vulnerable sequence. Analysis of the cleaved increase in fluorescence correlates with the presence of cancer in the tissue. The contrast agent and fluorescent microscopy gives an initial estimate of residual cancer on the surface of the tissue. However, because evaluation by fluorescence microscopy alone can be subject to non-specific positive margin highlighting, the utility of adding OCT and RCM becomes crucial in margin analysis. The RCM provides submicron detail to inform tissue-type classification, and guide OCT data segmentation, which provides volumetric mapping of the extent of the tumor infiltration. By using the FL mode to identify areas of highly suspicious for presence of cancer, the acquisition time can be greatly reduced during RCM/OCT imaging, which is critical for intraoperative utility. The information from the three modalities was combined in an image-processing regimen to automatically detect the cancer margins in 3D very effectively.

Five cases were highlighted in this paper to demonstrate the potential clinical utility of the multimodal optical imaging for breast cancer surgery guidance. The first case presented showed uniform staining with a near complete increase in FL signal from the contrast agent. RCM/OCT imaging and subsequent image analysis showed similar agreement, that the margins extended to the surface and edges and that the tumor content was distributed throughout the depth. Histopathological analysis confirmed that the specimen was an invasive ductal carcinoma, and the annotated section correlated well with the RCM/OCT images.

The second case was a more complicated specimen and more representative of the type of tissue found at tumor boundaries. Visual inspection showed a clear interface between yellow, likely fat and whiter, potentially malignant tissue. The FL image showed increased brightness from the contrast agent in approximately the same areas as observed visually, though the signal ratio was not as high as in case 1. The RCM/OCT imaging showed a more complicated tissue structure than was seen in case 1. The annotated histology of the hematoxylin and eosin stained tissue section indicated invasive ductal carcinoma amidst normal breast tissue. At certain depths, the histological annotations matched well with the image analysis prediction. However, through the analysis of the volumetric OCT data, we predict the tumor extended laterally in depth. In this study, the specimens were sectioned only a limited number of times. For 3D ground truth from histology, the specimen would need to be sectioned serially throughout, which is costly, introduces delays, and requires lengthy expert interpretation. This emphasizes that further improvements of this instrument is feasible which will be useful for eventual incorporation into surgical pathology practice.

The third presented case is one that visually appears like uniform adipose tissue. The FL channel did not show increased fluorescence signal and the histopathologist indicated no tumor content. Although in real clinical practice additional OCT imaging would be unnecessary when RCM indicates the absence of the positive margins, in this case we performed OCT imaging to confirm that the segmentation algorithm assigns this area to the correct tissue class.

The fourth case is that of a heterogeneous specimen, where multiple tissue types can be visualized by eye. The fluorescent contrast agent showed a small region of potentially positive margins, which warranted more in-depth analysis. However, histological annotation showed the presence of only multiple benign breast lobules and not of cancer. OCT also showed some small false positives and the lobules areas. However, RCM confirmed the presence of benign tissue.

The fifth case showed the utility of OCT depth sectioning capabilities, as hidden positive margin areas can subside underneath the specimen surface, within 1 mm depth. Histology might fail showing these areas if, as usually in the clinical practice, a single slice at a single depth is analyzed.

Histology correlation of the imaging findings was difficult, as perfect correlation of the OCT/RCM images with histology was almost impossible due to alterations of tissue alignment with processing of the tissue during histology preparation. This is particularly apparent in samples with a high content of adipose tissue. For this reason, precise percentages of tissue type area coverage were not computed or compared. Furthermore, the histological practice often outlines large areas that have a mix of tissues within them out of an abundance of caution and to ensure that no tumor cells remain. For example, fat cells within a tumor area, would be segmented as fat cells even though they would be located in a region marked for removal.

The segmentation algorithm correctly classified many complex tissue types as benign, such as fibrosis and treatment-related change from malignant to benign. However, some false positives after OCT image segmentation were still present in two cases. As RCM cannot be performed fast enough on large areas highlighted by FL imaging, this technology needs further improvements to eliminate the false positives. Potentially, larger training sets may improve the false positive rate in areas of normal lobules, as well as the use of more sophisticated algorithms for tissue-type differentiation. More precise histology alignment and preparation will potentially allow for registration of images and tissue type local area comparisons.

Taken together, this case study suggests the novel multimodal instrument can be of great use in identifying surgical margins on breast cancer resection specimens. As noted, rapid feedback of margin extent and location could reduce the need for repeated surgeries and cosmetic damage.

5. Conclusion

Multimodal optical imaging was performed on breast surgical specimens with the goal of evaluating the potential of this technology for assessing surgical margins as a tool for surgery guidance. Twenty specimens were evaluated, and a reasonably good agreement between imaging findings and histopathology findings were not noted. However, overestimation of positive margins was noted on several tissue specimens as compared to histology findings. Although some of the overestimates might indeed be false-positives, it is also possible that they were true-positives, considering perfect histology correlation was impossible due to registration obstacles. Some small false positives were flagged by OCT algorithmic segmentation in areas of normal lobules. RCM analysis correctly rejected these false positives in most of the cases. Histology processing usually distorts the shape of the specimen, particularly in tissues high in adipocytes. In addition, a single histology slide was taken from each surface of the surgical specimen from a depth of approximately 100 um beneath the specimen surface, while RCM and OCT imaging was performed to depths of 100 μm and 1mm, respectively. The deeper, volumetric information sometimes revealed margins that extended beyond the histology indication. Further evaluation of this technology on additional specimens and better correlation with histology is ongoing which will be reported in a follow-up study. Nevertheless, the preliminary findings are encouraging, showing the potential benefits of this technology for intraoperative evaluation of breast surgical specimens.

6. Limitations and future work

Despite encouraging preliminary results, improvements on instrumentation, data collection and data analysis will be considered for future studies. Registration of OCT/RCM images and histological must be reconsidered. The specimens here were photographed and marked with a dot of tissue dye to aid in alignment for sectioning. However, there are undoubtedly slight misalignments in pitch and roll rotations. This affects the accuracy of the matching tumor percentages globally and spatially between RCM/OCT and histology. Future studies will attempt to utilize custom cassettes during fixation or imaging and sectioning will be performed using frozen sections within the same laboratory space as soon as possible for better correlation of histology and imaging. The processing algorithm will also need to be based on larger training sets, as there is significant variation within the scattering properties of tissue specimens belonging to the same tissue class, which contributes to the false-positive rate in normal lobule areas.

Further studies will also attempt to analyze the contrast agent in more details in examine potential options for providing more selective cancer binding, as well as to add a fluorescence channel to the confocal system, as fluorescence CM shall improve pathologist ability to differentiate tissue-types, as demonstrated by other studies [31].

This paper presents only limited case studies; further evaluation of the instrument will be performed using more resection specimens. This will greatly improve the image processing and automated classification techniques. As datasets become larger and more diverse, machine learning techniques can be implemented.

Data Availability

All relevant data are within the manuscript.

Funding Statement

Nicusor Iftimia-PI 2R44CA173998-02A1 US National Cancer Institute The funder provided support in the form of salaries for authors [MS, SK, GM, DP, JP, JG, MS, KB, NI], but did not have any additional role in the study design, data collection and analysis, decision to publish, or preparation of the manuscript. The specific roles of these authors are articulated in the ‘author contributions’ section.

References

  • 1.Azu M, Abrahamse P, Katz SJ, Jagsi R, Morrow M. What is an Adequate Margin for Breast-Conserving Surgery? Surgeon Attitudes and Correlates. Ann Surg Oncol. 2010. February;17(2):558–63. 10.1245/s10434-009-0765-1 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Blair SL, Thompson K, Rococco J, Malcarne V, Beitsch PD, Ollila DW. Attaining negative margins in breast-conservation operations: is there a consensus among breast surgeons? J Am Coll Surg. 2009. November;209(5):608–13. 10.1016/j.jamcollsurg.2009.07.026 [DOI] [PubMed] [Google Scholar]
  • 3.Morrow M. Breast conservation and negative margins: how much is enough? The Breast. 2009. October 1;18:S84–6. 10.1016/S0960-9776(09)70279-6 [DOI] [PubMed] [Google Scholar]
  • 4.Meric F, Mirza NQ, Vlastos G, Buchholz TA, Kuerer HM, Babiera GV, et al. Positive surgical margins and ipsilateral breast tumor recurrence predict disease-specific survival after breast-conserving therapy. Cancer. 2003. February 15;97(4):926–33. 10.1002/cncr.11222 [DOI] [PubMed] [Google Scholar]
  • 5.Swanson GP, Rynearson K, Symmonds R. Significance of margins of excision on breast cancer recurrence. Am J Clin Oncol. 2002. October;25(5):438–41. 10.1097/00000421-200210000-00002 [DOI] [PubMed] [Google Scholar]
  • 6.Fleming FJ, Hill ADK, Mc Dermott EW, O’Doherty A, O’Higgins NJ, Quinn CM. Intraoperative margin assessment and re-excision rate in breast conserving surgery. Eur J Surg Oncol J Eur Soc Surg Oncol Br Assoc Surg Oncol. 2004. April;30(3):233–7. 10.1016/j.ejso.2003.11.008 [DOI] [PubMed] [Google Scholar]
  • 7.Waljee JF, Hu ES, Newman LA, Alderman AK. Predictors of Re-excision among Women Undergoing Breast-Conserving Surgery for Cancer. Ann Surg Oncol. 2008. February 8;15(5):1297 10.1245/s10434-007-9777-x [DOI] [PubMed] [Google Scholar]
  • 8.Méndez JE, Lamorte WW, de las Morenas A, Cerda S, Pistey R, King T, et al. Influence of breast cancer margin assessment method on the rates of positive margins and residual carcinoma. Am J Surg. 2006. October 1;192(4):538–40. 10.1016/j.amjsurg.2006.06.009 [DOI] [PubMed] [Google Scholar]
  • 9.McLaughlin SA, Ochoa-Frongia LM, Patil SM, Cody HS, Sclafani LM. Influence of frozen-section analysis of sentinel lymph node and lumpectomy margin status on reoperation rates in patients undergoing breast-conservation therapy. J Am Coll Surg. 2008. January;206(1):76–82. 10.1016/j.jamcollsurg.2007.07.021 [DOI] [PubMed] [Google Scholar]
  • 10.Menes TS, Tartter PI, Mizrachi H, Smith SR, Estabrook A. Touch preparation or frozen section for intraoperative detection of sentinel lymph node metastases from breast cancer. Ann Surg Oncol. 2003. December;10(10):1166–70. 10.1245/aso.2003.04.023 [DOI] [PubMed] [Google Scholar]
  • 11.Cox CE, Ku NN, Reintgen DS, Greenberg HM, Nicosia SV, Wangensteen S. Touch Preparation Cytology of Breast Lumpectomy Margins With Histologic Correlation. Arch Surg. 1991. April 1;126(4):490–3. 10.1001/archsurg.1991.01410280094014 [DOI] [PubMed] [Google Scholar]
  • 12.Jacobs L. Positive margins: the challenge continues for breast surgeons. Ann Surg Oncol. 2008. May;15(5):1271–2. 10.1245/s10434-007-9766-0 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Cendán JC, Coco D, Copeland EM. Accuracy of Intraoperative Frozen-Section Analysis of Breast Cancer Lumpectomy-Bed Margins. J Am Coll Surg. 2005. August 1;201(2):194–8. 10.1016/j.jamcollsurg.2005.03.014 [DOI] [PubMed] [Google Scholar]
  • 14.McClatchy DM, Rizzo EJ, Meganck J, Kempner J, Vicory J, Wells WA, et al. Calibration and analysis of a multimodal micro-CT and structured light imaging system for the evaluation of excised breast tissue. Phys Med Biol. 2017. November 10;62(23):8983–9000. 10.1088/1361-6560/aa94b6 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Qiu S-Q, Dorrius MD, de Jongh SJ, Jansen L, de Vries J, Schröder CP, et al. Micro-computed tomography (micro-CT) for intraoperative surgical margin assessment of breast cancer: A feasibility study in breast conserving surgery. Eur J Surg Oncol J Eur Soc Surg Oncol Br Assoc Surg Oncol. 2018;44(11):1708–13. 10.1016/j.ejso.2018.06.022 [DOI] [PubMed] [Google Scholar]
  • 16.Tan JE, Orel SG, Schnall MD, Schultz DJ, Solin LJ. Role of magnetic resonance imaging and magnetic resonance imaging—guided surgery in the evaluation of patients with early-stage breast cancer for breast conservation treatment. Am J Clin Oncol. 1999. August;22(4):414–8. 10.1097/00000421-199908000-00020 [DOI] [PubMed] [Google Scholar]
  • 17.Moschetta M, Telegrafo M, Introna T, Coi L, Rella L, Ranieri V, et al. Role of specimen US for predicting resection margin status in breast conserving therapy. Il G Chir. 2015. Oct;36(5):201–4. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Kalinyak JE, Berg WA, Schilling K, Madsen KS, Narayanan D, Tartar M. Breast cancer detection using high-resolution breast PET compared to whole-body PET or PET/CT. Eur J Nucl Med Mol Imaging. 2014. February;41(2):260–75. 10.1007/s00259-013-2553-1 [DOI] [PubMed] [Google Scholar]
  • 19.Narayanan D, Madsen KS, Kalinyak JE, Berg WA. Interpretation of positron emission mammography and MRI by experienced breast imaging radiologists: performance and observer reproducibility. AJR Am J Roentgenol. 2011. April;196(4):971–81. 10.2214/AJR.10.5081 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Thill M, Röder K, Diedrich K, Dittmer C. Intraoperative assessment of surgical margins during breast conserving surgery of ductal carcinoma in situ by use of radiofrequency spectroscopy. Breast Edinb Scotl. 2011. December;20(6):579–80. 10.1016/j.breast.2011.08.134 [DOI] [PubMed] [Google Scholar]
  • 21.Troyan SL, Kianzad V, Gibbs-Strauss SL, Gioux S, Matsui A, Oketokoun R, et al. The FLARE intraoperative near-infrared fluorescence imaging system: a first-in-human clinical trial in breast cancer sentinel lymph node mapping. Ann Surg Oncol. 2009. October;16(10):2943–52. 10.1245/s10434-009-0594-2 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Kennedy KM, Chin L, McLaughlin RA, Latham B, Saunders CM, Sampson DD, et al. Quantitative micro-elastography: imaging of tissue elasticity using compression optical coherence elastography. Sci Rep. 2015. October 27;5:15538 10.1038/srep15538 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Sharma V, Shivalingaiah S, Peng Y, Euhus D, Gryczynski Z, Liu H. Auto-fluorescence lifetime and light reflectance spectroscopy for breast cancer diagnosis: potential tools for intraoperative margin detection. Biomed Opt Express. 2012. July 9;3(8):1825–40. 10.1364/BOE.3.001825 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Weyers B, Marsden M, Sun T, Bec J, Bewley AF, Gandour-Edwards RF, et al. Fluorescence lifetime imaging for intraoperative cancer delineation in transoral robotic surgery—Weyers— 2019. —Translational Biophotonics—Wiley Online Library; [Internet]. [cited 2020 Aug 14]. Available from: https://onlinelibrary.wiley.com/doi/full/10.1002/tbio.201900017 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.Nguyen FT, Zysk AM, Chaney EJ, Kotynek JG, Oliphant UJ, Bellafiore FJ, et al. Intraoperative Evaluation of Breast Tumor Margins with Optical Coherence Tomography. Cancer Res. 2009. November 15;69(22):8790–6. 10.1158/0008-5472.CAN-08-4340 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Savastru D, Chang EW, Miclos S, Pitman MB, Patel A, Iftimia N. Detection of breast surgical margins with optical coherence tomography imaging: a concept evaluation study. J Biomed Opt. 2014. May;19(5):056001 10.1117/1.JBO.19.5.056001 [DOI] [PubMed] [Google Scholar]
  • 27.Thomas G, Nguyen T-Q, Pence IJ, Caldwell B, O’Connor ME, Giltnane J, et al. Evaluating feasibility of an automated 3-dimensional scanner using Raman spectroscopy for intraoperative breast margin assessment. Sci Rep. 2017. 19;7(1):13548 10.1038/s41598-017-13237-y [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Piras D, Xia W, Steenbergen W, van Leeuwen TonG, Manohar S. Photoacoustic Imaging of the Breast Using the Twente Photoacoustic Mammoscope: Present Status and Future Perspectives. IEEE J Sel Top Quantum Electron. 2010. July;16(4):730–9. [Google Scholar]
  • 29.Maloney BW, McClatchy DM, Pogue BW, Paulsen KD, Wells WA, Barth RJ. Review of methods for intraoperative margin detection for breast conserving surgery. J Biomed Opt. 2018;23(10):1–19. 10.1117/1.JBO.23.10.100901 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30.Zhang W, Ling D, Tan J, Zhang J, Li L. Expression of urokinase plasminogen activator and plasminogen activator inhibitor type-1 in ovarian cancer and its clinical significance. Oncol Rep. 2013. February 1;29(2):637–45. 10.3892/or.2012.2148 [DOI] [PubMed] [Google Scholar]
  • 31.Krishnamurthy S, Brown JQ, Iftimia N, Levenson RM, Rajadhyaksha M. Ex Vivo Microscopy A Promising Next-Generation Digital Microscopy Tool for Surgical Pathology Practice. Arch Pathol Lab Med. 2019;143:1058–68 10.5858/arpa.2019-0058-RA [DOI] [PMC free article] [PubMed] [Google Scholar]

Decision Letter 0

Matthew Bogyo

15 Oct 2020

PONE-D-20-28159

Assessment of breast cancer surgical margins with multimodal optical microscopy: a preliminary clinical study

PLOS ONE

Dear Dr. Iftimia,

Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.

As you will see the two reviewers have some significant issues with the manuscript that need to be addressed. Specifically, the first reviewer brings up a number of points about lack of data on the target specificity, kinetics of probe activation, depth penetrance and analytical characterization. These are concerns that need to be addressed in your revision. The second reviewer mainly had minor comments and questions that are marked up in the pdf file attached. Please address all of these comments as well. I am happy to consider a revision of the work as long as you are able to address all of the reviewer comments with the addition of new data or with a reasonable explanation as to why additional data is not necessary.

Please submit your revised manuscript by Nov 29 2020 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.

Please include the following items when submitting your revised manuscript:

  • A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'.

  • A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'.

  • An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'.

If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter.

If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: http://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols

We look forward to receiving your revised manuscript.

Kind regards,

Matthew Bogyo, Ph.D.

Academic Editor

PLOS ONE

Journal Requirements:

When submitting your revision, we need you to address these additional requirements.

1. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at

https://journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and

https://journals.plos.org/plosone/s/file?id=ba62/PLOSOne_formatting_sample_title_authors_affiliations.pdf

2. Please ensure your Methods and reagents are be described in sufficient detail for another researcher to reproduce the experiments described. Specifically, please provide further details on the methods of section 2.1 Flourescence imaging agent, including details of conditions for synthesis.

3. Please include the results for the other 10 samples as supplementary information.

4. Please upload your figures as separate files.

5. In the ethics statement in the manuscript and in the online submission form, please provide additional information about the patient samples used in your retrospective study, including: a) the date range (month and year) during which patients' samples records were retrieved; b) the date range (month and year) during which patients whose samples were selected for this study sought treatment. If patients provided informed written consent to have samples from their medical records used in research, please include this information.

6. Thank you for including your ethics statement: 'Breast surgical specimens were provided by MD Anderson Cancer Center. Fourteen deidentified specimens were acquired from breast cancer patients undergoing surgery procedures (both lumpectomy and mastectomy) under the IRB approved protocol PA14-1036.'

a. Please amend your current ethics statement to include the full name of the ethics committee/institutional review board(s) that approved your specific study and confirm that your named institutional review board or ethics committee specifically approved this study.

b. Once you have amended this/these statement(s) in the Methods section of the manuscript, please add the same text to the “Ethics Statement” field of the submission form (via “Edit Submission”).

For additional information about PLOS ONE ethical requirements for human subjects research, please refer to http://journals.plos.org/plosone/s/submission-guidelines#loc-human-subjects-research

7.Thank you for stating the following in the Financial Disclosure section:

[Nicusor Iftimia-PI

2R44CA173998-02A1

US National Cancer Institute

The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript.].   

We note that one or more of the authors are employed by a commercial company: Physical Sciences Inc

  1. Please provide an amended Funding Statement declaring this commercial affiliation, as well as a statement regarding the Role of Funders in your study. If the funding organization did not play a role in the study design, data collection and analysis, decision to publish, or preparation of the manuscript and only provided financial support in the form of authors' salaries and/or research materials, please review your statements relating to the author contributions, and ensure you have specifically and accurately indicated the role(s) that these authors had in your study. You can update author roles in the Author Contributions section of the online submission form.

Please also include the following statement within your amended Funding Statement.

“The funder provided support in the form of salaries for authors [insert relevant initials], but did not have any additional role in the study design, data collection and analysis, decision to publish, or preparation of the manuscript. The specific roles of these authors are articulated in the ‘author contributions’ section.”

If your commercial affiliation did play a role in your study, please state and explain this role within your updated Funding Statement.

2. Please also provide an updated Competing Interests Statement declaring this commercial affiliation along with any other relevant declarations relating to employment, consultancy, patents, products in development, or marketed products, etc.  

Within your Competing Interests Statement, please confirm that this commercial affiliation does not alter your adherence to all PLOS ONE policies on sharing data and materials by including the following statement: "This does not alter our adherence to  PLOS ONE policies on sharing data and materials.” (as detailed online in our guide for authors http://journals.plos.org/plosone/s/competing-interests) . If this adherence statement is not accurate and  there are restrictions on sharing of data and/or materials, please state these. Please note that we cannot proceed with consideration of your article until this information has been declared.

Please include both an updated Funding Statement and Competing Interests Statement in your cover letter. We will change the online submission form on your behalf.

Please know it is PLOS ONE policy for corresponding authors to declare, on behalf of all authors, all potential competing interests for the purposes of transparency. PLOS defines a competing interest as anything that interferes with, or could reasonably be perceived as interfering with, the full and objective presentation, peer review, editorial decision-making, or publication of research or non-research articles submitted to one of the journals. Competing interests can be financial or non-financial, professional, or personal. Competing interests can arise in relationship to an organization or another person. Please follow this link to our website for more details on competing interests: http://journals.plos.org/plosone/s/competing-interests

[Note: HTML markup is below. Please do not edit.]

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #1: No

Reviewer #2: Yes

**********

2. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #1: Yes

Reviewer #2: No

**********

3. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #1: No

Reviewer #2: Yes

**********

4. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #1: Yes

Reviewer #2: Yes

**********

5. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #1: There are several aspects of this manuscript that are very exciting and have the potential to impact medical management of BCS. This manuscript presents data to demonstrate the utility of the method. However, the manuscript does not present enough information about aspects of the technology to fully interpret its impact. I will review the data presented and then the overall impact of the data.

Data presented:

Overall the authors are assessing a multi-modality optical imaging modality to assess tumor margins in lumpectomy tissues ex vivo. The approach has been developed to identify regions likely to have cancer using FL guided imaging (and a probe to assess tumor presence) and then utilizes other optical approaches to verify tumor presence and depth of tumor lesions (OCT and RCM).

Probe: The description of the probe defines it as a Cy5.5 labeled probe that is self-quenched. The quench is released by uPA protease. The probe is also targeted to Her2 receptors, the sequence of the targeting peptide is withheld from the readers. There is in vitro data to describe the dequenching of probe and the level of activation. Methodology of use is quite straight forward. The probe is simply applied to ex vivo tumor resection samples for 1-2 minutes, rinsed and then FL measured. If FL regions are detected other means of imaging are used to confirm presence of cancer and report on its penetration. The level of un-quenched probe washout after a 2-minute incubation and washing should be assessed. Perhaps specific signal is lost.

There is no data to demonstrate specificity of the probe for Her2 binding, nor protease activation. It would be good for there to be data explaining the tissue characteristics of probe activation. Using protease inhibitors specificity of probe activation could be established. Similarly using peptide competitors for Her2, receptor binding could be assessed. What is the kinetics for Her2 binding to the peptide in tissues samples, and for that matter do in vitro protease activation kinetics represent probe activation kinetics in tissue samples. Using mouse models with topical application of the probe, these could easily be answered.

It is not clear why the probe should penetrate tissue at all given the short duration of incubation. Solubility measures of the probe should be presented and penetration studies into human tumors grown in mice would help inform the statement of 1 mm probe penetration. In general, topical applications of the probe is aqueous carriers do not promote tissue penetration.

Finally, a reasonable discussion of the expression level of uPA in breast cancer tumors should be placed within the paper and include the presence of Her2 receptor as well. Justification for the synthetic design of this probe is sorely needed.

Patient samples: Given the Her2 binding ability of the probe the patient samples used in the study should be identified as being Her2 positive or not. This information is not provided. Her2 only is overexpressed in around 20% of breast cancer patients, so it is not clear if utilizing the Her2 peptide ligand is helpful or not. Her2-positive and Her2-negative samples could be compared to describe the impact of the peptide ligand on tumor tissue recognition.

The paper starts off claiming 100% correlation of the data with ground truth histology but later reveals that the correlation is only in histology verified positive lumpectomy specimens, and that false positives were detected. It is not clear if the multimodality approach was able to resolve these before or after histological verification.

Cancer visualization: It is not clear the sensitivity of the methodology. Many pathological assessments call positive margins with very few cancer cells present. What is the lowest number of cancer cells that can be FL detected? The time for such analysis also needs to be demonstrated. The author states it speeds up analysis, but this is due to reducing the amount of tissue that OCT or Raman are used to study. What is the total average time to sample and analyze the lumpectomy samples for the presences of cancer? Is it faster than frozen sectioning?

The author also states that correlations of FL and OCT images back to fixed tissue is difficult or impossible to perform. This has been solved for skin cancer and the author is referred to Walker et al., Cancer Research 2020.

It is not clear is the cancer specimens were assessed from all sides. To gain true assessment of the utility of this approach all edges and sides of the tumor specimens must be assessed. With that in mind it would be good for the author to present sensitivity and specificity table for the samples analyzed.

Was the study blinded to histology outcomes?

Overall Impact of the Study:

The impact of the approaches demonstrated here could be significant for management of breast conserving surgery. However, many of the parameters that are required for this assessment are missing, e.g. specificity of the probe, Her2 positivity of tested samples, time to perform the procedure, the ability to “see” the entire surface of the lump and stitch together a useful map of cancer at the margin to go back to the surgical cavity and in real time re-excise regions with positive margins. This would have been interesting to discuss in the future perspectives of the discussion section.

Reviewer #2: It would be helpful to implement the attached edits within the pdf and have the manuscript read by a native English speaker.

Overall this a well-conducted study that suffers a bit from a small sample size but highlights how the sum of multiple imaging modalities can make up for deficiencies inherent in any single method.

**********

6. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #1: No

Reviewer #2: Yes: Jonathan Sorger

[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.]

While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step.

Attachment

Submitted filename: PONE-D-20-28159.pdf

PLoS One. 2021 Feb 11;16(2):e0245334. doi: 10.1371/journal.pone.0245334.r002

Author response to Decision Letter 0


1 Dec 2020

Reviewer 1:

Q1: The level of un-quenched probe washout after a 2-minute incubation and washing should be assessed. Perhaps specific signal is lost.

Answer: The unquenched probe contributes to the background fluorescence signal, as observed from Fig. 2 (fluorescence background before cleavage- created by the unquenched agent). This level is rather small, as the fluorescence increases 10 fold after peptide cleavage. However, we believe that the reviewer refers to the un-cleaved peptide after 2 minutes of incubation (see plot from Fig. 3 - revised manuscript). This level can vary as a function of the enzyme-substrate ratio. The more enzyme on the tissue, the faster is the cleavage -see E:S = 1:25 vs. E:S= 1:200. We noticed a similar behavior during the experiments performed on tissue specimens. Since not all specimens express the same level of uPA enzyme, the level of un-cleaved peptide varies after 2-minute of incubation. To determine the amount of un-cleaved peptide left after incubation, we added enzyme to the washout solution from several specimens and noticed no fluorescence increase in some cases and up to 20% fluorescence increase, confirming that the uPA concentration varies from one specimen to another. For specimens that express lower uPA levels, longer incubation provides improved cleavage. However, since small concentrations of uPA are present as well in the surrounding benign fibrotic tissue, longer incubation times actually result in a decrease of the fluorescence contrast of the tumor tissue. Based on these facts, we determined that 2 minutes of incubation provides optimal fluorescence contrast.

Q 2. There is no data to demonstrate specificity of the probe for Her2 binding, nor protease activation. It would be good for there to be data explaining the tissue characteristics of probe activation. Using protease inhibitors specificity of probe activation could be established. Similarly using peptide competitors for Her2, receptor binding could be assessed. What is the kinetics for Her2 binding to the peptide in tissues samples, and for that matter do in vitro protease activation kinetics represent probe activation kinetics in tissue samples. Using mouse models with topical application of the probe, these could easily be answered.

Answer: Unfortunately, our funded NIH grant does not support animal studies, and therefore this experiment cannot be performed as suggested. Furthermore, it was not the focus of this research to study the effectiveness of the contrast agent in tagging the Her2 receptor. This would warrant a full separate study, as it was not the intent of our current study. Therefore, we removed from the manuscript any discussion related to this subject. We have changed the manuscript as follows: “A spacer (aminohexanoic acid) may be used to ensure that this peptide moiety will maintain fast uPA cleavage kinetics. However, enhancement of cancer specificity was not a major focus of this preliminary study.”

Q3. It is not clear why the probe should penetrate tissue at all given the short duration of incubation. Solubility measures of the probe should be presented and penetration studies into human tumors grown in mice would help inform the statement of 1 mm probe penetration. In general, topical applications of the probe is aqueous carriers do not promote tissue penetration.

Answer: We agree with the reviewer that 2 minutes of incubation might not provide very deep tissue penetration. However, 1 mm penetration is sufficient, as at least 1 mm safe margin is needed in surgery. Based on our experiments, after 2 minutes of contrast agent placement on the top surface (pipette drops), the opposite side on a specimen with 1 mm thickness showed similar fluorescence on both surfaces. However, as suggested, further experiments will be performed in a subsequent study to address further increase in solubility measures.

Therefore, the manuscript was modified as follows: “This suggests that that the optimum incubation time will be about 1 minute. However, to enable contrast agent diffusion in the tissue, our experiments showed that a maximum of 2 minutes incubation time is a good compromise between fluorescence contrast and agent diffusivity to about 1 mm deep. For longer incubation times, the smaller concentrations of the uPA enzyme present on the benign tissue start cleaving the agent, and thus degrade tumor contrast.”

Q4. Finally, a reasonable discussion of the expression level of uPA in breast cancer tumors should be placed within the paper and include the presence of Her2 receptor as well. Justification for the synthetic design of this probe is sorely needed.

Answer: The expression level of the uPA in cancer tumors is discussed in the paper: "Based on literature data, we estimated the concentration of the enzyme that is present in malignant tissue versus benign tissue, and used these numbers as a baseline for our measurements. Literature reports (30) show that the average uPA level in benign tissue is 1 ng/mg protein, while in tumor tissue is 8 ng/mg protein." The synthetic design of the probe is not the main goal of this paper, as in our study we focused on the microscopy aspects. Any other contrast agents can be used with this microscope. Therefore, we modified the manuscript as follows: “ A contrast agent capable of highlighting areas of the specimen that have potential positive margins can be used with the proposed microscope. The main requirement is to have close to 100% sensitivity, as otherwise positive margins may be left behind. For this study, a novel contrast agent was developed by our team. It consists of an optically silent peptide substrate containing two (near-infrared; NIR) fluorochromes (Cy5.5), internally quenched. This agent was synthesized to track positive margins on breast lumpectomy specimens via specific urokinase plasminogen activator (uPA) cleavage.”

Q5. Patient samples: Given the Her2 binding ability of the probe the patient samples used in the study should be identified as being Her2 positive or not. This information is not provided. Her2 only is overexpressed in around 20% of breast cancer patients, so it is not clear if utilizing the Her2 peptide ligand is helpful or not. Her2-positive and Her2-negative samples could be compared to describe the impact of the peptide ligand on tumor tissue recognition.

Answer: Her-2 positive and Her-2 negative specimens cannot be easily identified and thus such experiments were beyond the scope of this paper. HER2 testing through immunohistochemistry and FISH has shown error rates as high as 20%. A key aspect of the presented technology is that other contrast agents may be used if necessary or preferred.

Q6. The paper starts off claiming 100% correlation of the data with ground truth histology but later reveals that the correlation is only in histology verified positive lumpectomy specimens, and that false positives were detected. It is not clear if the multimodality approach was able to resolve these before or after histological verification.

Answer: From the presented data it is clear that the assessment of multimodal imaging effectiveness was performed by comparing imaging findings against histopathology findings. The imaging study was first performed and then the imaging results were compared against histopathology results. A table was inserted to show the summary of the findings.

Table 1: Summary of imaging and histopathology findings

20 Specimen study Fluorescence imaging RCM guided OCT segmentation Histopathology

Positive margins specimens 12 10 (2FP) 8 TP

Negative margins specimens 8 10 (0FN) 12 TN

RCM-guided OCT predictive values: PPV=1.0; NPV= 0.83

Q7. Cancer visualization: It is not clear the sensitivity of the methodology. Many pathological assessments call positive margins with very few cancer cells present. What is the lowest number of cancer cells that can be FL detected? The time for such analysis also needs to be demonstrated. The author states it speeds up analysis, but this is due to reducing the amount of tissue that OCT or Raman are used to study. What is the total average time to sample and analyze the lumpectomy samples for the presences of cancer? Is it faster than frozen sectioning?

Answer: Epifluorescence imaging cannot be used visualize individual cancer cells, as its resolution is ~ 50 um. Therefore, only clusters of cells can be visualized. Single cancer cell detection is out of the question, and not really important, as the usual clinical protocol is to have a boost radiation of the surgical bed and destroy eventual small clusters of cells. Indeed, fluorescence imaging reduces the time for microscopy analysis. The analysis time varies as a function of the size of the fluorescence-indicated margin. In our study, this area varied between 25 mm2 and 200 mm2. With an OCT frame rate of 40 frames/sec (40 B-scans/sec), a B-scan size of 2.2 mm, and a 20 um separation between B-scans, an area of 100 mm2 requires ~2 minutes. Again, frozen sectioning is not suitable for breast specimens, as they have a high adipose content and are thus very difficult to section.

Q8. The author also states that correlations of FL and OCT images back to fixed tissue is difficult or impossible to perform. This has been solved for skin cancer and the author is referred to Walker et al., Cancer Research 2020.

Answer: The imaged specimens were marked with ink on the top surface and histology processing was carefully performed to correlate the correct side with imaging results. However, when we refer to imaging-histology correlation, we consider the fact that when fixing the tissue, it shrinks and thus the histology images of the specimen do not look exactly like the enface OCT images, and thus 100% correlation of the findings is not really possible. Furthermore, skin cancer and epidermal/dermal tissue maintain fidelity during sectioning much better than tissues high in adipose. Difficulties surrounding sectioning of adipose tissue include: adipose tissue is more prone to ripping, requires lower temperatures for frozen sectioning, and requires longer delay times for embedding (in paraffin or OCT) to ensure the embedding material supports the structure.

Q9. It is not clear is the cancer specimens were assessed from all sides. To gain true assessment of the utility of this approach all edges and sides of the tumor specimens must be assessed. With that in mind it would be good for the author to present sensitivity and specificity table for the samples analyzed.

R: The tissue specimens were relatively thin (~ 3-4 mm when flattened within the instrument specimen holder). Therefore, they were analyzed on the two opposite surfaces. A table showing the positive and negative predictive values from a limited set of specimens was inserted within the revised manuscript. Since this was a feasibility study on a small set of specimens, a clear assessment of technology sensitivity and specificity might be premature.

Q10. Was the study blinded to histology outcomes?

Answer: Yes, the microscopy data were first analyzed and then tested against the histopathology reports.

Q11. Overall Impact of the Study:

The impact of the approaches demonstrated here could be significant for management of breast conserving surgery. However, many of the parameters that are required for this assessment are missing, e.g. specificity of the probe, Her2 positivity of tested samples, time to perform the procedure, the ability to “see” the entire surface of the lump and stitch together a useful map of cancer at the margin to go back to the surgical cavity and in real time re-excise regions with positive margins. This would have been interesting to discuss in the future perspectives of the discussion section.

Answer. We agree that a more in-depth study will be needed to further analyze these aspects. However the main goal of this paper was to report the potential impact of this technology. The conclusion section has been amended with the next paragraph: “Further studies will also attempt to analyze the contrast agent in more details in examine potential options for providing more selective cancer binding. This paper presents only limited case studies; further evaluation of the instrument will be performed using more resection specimens. This will greatly improve the image processing and automated classification techniques. As datasets become larger and more diverse, machine learning techniques can be implemented. ”

Reviewer 2:

1. It would be helpful to implement the attached edits within the pdf and have the manuscript read by a native English speaker.

2. Answer: English speaker proofreading was performed. We thank the reviewer for their careful proofreading and suggestions.

2. Overall this a well-conducted study that suffers a bit from a small sample size but highlights how the sum of multiple imaging modalities can make up for deficiencies inherent in any single method.

Answer: We agree with the reviewer that this was a relatively small sample size study. We included an additional of 6 cases and a table indicating technology performance.

Attachment

Submitted filename: R2R_28159R1.docx

Decision Letter 1

Matthew Bogyo

29 Dec 2020

Assessment of breast cancer surgical margins with multimodal optical microscopy: a feasibility clinical study

PONE-D-20-28159R1

Dear Dr. Iftimia,

We’re pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it meets all outstanding technical requirements.

Within one week, you’ll receive an e-mail detailing the required amendments. When these have been addressed, you’ll receive a formal acceptance letter and your manuscript will be scheduled for publication.

An invoice for payment will follow shortly after the formal acceptance. To ensure an efficient process, please log into Editorial Manager at http://www.editorialmanager.com/pone/, click the 'Update My Information' link at the top of the page, and double check that your user information is up-to-date. If you have any billing related questions, please contact our Author Billing department directly at authorbilling@plos.org.

If your institution or institutions have a press office, please notify them about your upcoming paper to help maximize its impact. If they’ll be preparing press materials, please inform our press team as soon as possible -- no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org.

Kind regards,

Matthew Bogyo, Ph.D.

Academic Editor

PLOS ONE

Additional Editor Comments (optional):

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation.

Reviewer #1: All comments have been addressed

Reviewer #2: All comments have been addressed

**********

2. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #1: Yes

Reviewer #2: Yes

**********

3. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #1: Yes

Reviewer #2: N/A

**********

4. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #1: Yes

Reviewer #2: Yes

**********

5. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #1: Yes

Reviewer #2: Yes

**********

6. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #1: (No Response)

Reviewer #2: (No Response)

**********

7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #1: No

Reviewer #2: No

Acceptance letter

Matthew Bogyo

29 Jan 2021

PONE-D-20-28159R1

Assessment of breast cancer surgical margins with multimodal optical microscopy: A feasibility clinical study

Dear Dr. Iftimia:

I'm pleased to inform you that your manuscript has been deemed suitable for publication in PLOS ONE. Congratulations! Your manuscript is now with our production department.

If your institution or institutions have a press office, please let them know about your upcoming paper now to help maximize its impact. If they'll be preparing press materials, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information please contact onepress@plos.org.

If we can help with anything else, please email us at plosone@plos.org.

Thank you for submitting your work to PLOS ONE and supporting open access.

Kind regards,

PLOS ONE Editorial Office Staff

on behalf of

Dr. Matthew Bogyo

Academic Editor

PLOS ONE

Associated Data

    This section collects any data citations, data availability statements, or supplementary materials included in this article.

    Supplementary Materials

    Attachment

    Submitted filename: PONE-D-20-28159.pdf

    Attachment

    Submitted filename: R2R_28159R1.docx

    Data Availability Statement

    All relevant data are within the manuscript.


    Articles from PLoS ONE are provided here courtesy of PLOS

    RESOURCES