Table 2.
Overall number of positive blood cultures obtained for different storage conditions, compared with no delay; and number of BacT/ALERT® bottle sensors that indicated growth prior to incubation
| Time (h) | Temp (°C) | Bottles (n) |
Positive (n)a | Positive (%) | P valuec | Sensor positived |
|---|---|---|---|---|---|---|
| 0 | NA | 90 | 80 | 88.9 | NA | 0/25 |
| 6 | 25 °C | 90 | 85 | 94.4 | 0.281 | 0/25 |
| 6 | Cool-box | 90 | 81 | 90.0 | 1.000 | 0/25 |
| 6 | 40 °C | 90 | 74 | 82.2 | 0.289 | 0/25 |
| 12 | 25 °C | 90 | 76 | 84.4 | 0.511 | 0/25 |
| 12 | Cool-box | 89b | 80 | 89.9 | 1.000 | 0/25 |
| 12 | 40 °C | 90 | 77 | 85.6 | 0.656 | 15/25 |
| 24 | 25 °C | 90 | 82 | 91.1 | 0.805 | 20/25 |
| 24 | Cool-box | 90 | 71 | 78.9 | 0.104 | 20/25 |
| 24 | 40 °C | 90 | 59 | 65.6 | < 0.001 | 20/25 |
| 6 | All | 270 | 240 | 88.9 | 1.000 | 0/75 |
| 12 | All | 269b | 233 | 86.6 | 0.787 | 15/75 |
| 24 | All | 270 | 212 | 78.5 | 0.030 | 60/75 |
| All | 25 °C | 270 | 243 | 90.0 | 0.841 | 20/75 |
| All | Cool-box | 269b | 232 | 86.2 | 0.694 | 20/75 |
| All | 40 °C | 270 | 210 | 77.8 | 0.021 | 35/75 |
aFor automated systems this is the number of bottles that flagged positive, and for manual systems this is the number of bottles that were positive on sub-culture; bOne bottle excluded due to contamination; cCompared with no delay; dSensor positive if BacT/ALERT® bottle sensor indicated growth prior to incubation. NA Not applicable