Skip to main content
. 2021 Feb 12;21:26. doi: 10.1186/s12880-021-00558-8

Table 2.

Bland–Altman statistics for the myocardial T1 measurements with MRmap as the reference method

Pre-contrast Post-contrast
Mean Difference, ms 95% Limits of Agreement, ms Mean Difference, ms 95% Limits of Agreement, ms
LM_python  − 0.24  − 5.11 to 4.63 0.72  − 5.81 to 7.25
RD_python  − 0.13  − 0.83 to 0.58 0.60  − 5.95 to 7.14
LM_C++_multi-core*  − 0.04  − 1.34 to 1.25 0.63  − 5.96 to 7.22
RD_C++_multi-core*  − 0.13  − 0.83 to 0.58 0.39  − 6.57 to 7.36

*T1 measurements between multi-core and single-core were the same