Table 3.
The percent difference from ADC reference value acquired on the 0.35 T MR‐LINAC at 0°C for all six vial concentrations and all gantry angles.
| Gantry angle (°) | Percent difference from concentration reference values (%) | |||||
|---|---|---|---|---|---|---|
| Conc. 1 | Conc. 2 | Conc. 3 | Conc. 4 | Conc. 5 | Conc. 6 | |
| 0 | ‐10.3% | ‐11.8% | ‐11.5% | ‐20.2% | ‐28.9% | 8.3% |
| 30 | ‐9.8% | ‐9.3% | ‐10.7% | ‐18.0% | ‐26.6% | ‐23.1% |
| 60 | ‐8.8% | ‐9.1% | ‐12.5% | ‐17.2% | ‐28.6% | ‐16.2% |
| 90 | ‐9.8% | ‐10.7% | ‐12.9% | ‐23.1% | ‐28.6% | ‐4.6% |
| 120 | ‐10.4% | ‐10.8% | ‐13.2% | ‐21.8% | ‐32.6% | ‐2.7% |
| 150 | ‐8.9% | ‐9.3% | ‐12.5% | ‐18.8% | ‐25.2% | 5.6% |
| 180 | ‐9.0% | ‐10.8% | ‐13.7% | ‐21.0% | ‐30.7% | ‐7.3% |
| 210 | ‐8.2% | ‐8.8% | ‐8.4% | ‐16.5% | ‐24.4% | 9.9% |
| 240 | ‐9.1% | ‐8.8% | ‐10.0% | ‐21.4% | ‐27.1% | ‐7.7% |
| 270 | ‐7.4% | ‐7.2% | ‐9.3% | ‐14.6% | ‐28.8% | ‐5.2% |
| 300 | ‐8.8% | ‐9.5% | ‐9.7% | ‐18.0% | ‐30.0% | ‐19.1% |
| 330 | ‐7.6% | ‐8.7% | ‐8.1% | ‐12.8% | ‐28.3% | ‐11.8% |