Skip to main content
. 2021 Jan 10;22(2):165–177. doi: 10.1002/acm2.13158

Table 4.

Results of FWHM in SPECT acquisition under each system. (a) 99mTc formulation; (b) 123I formulation.

FWHM Center X‐direction Y‐direction
Tangential Radial Tangential Radial Tangential Radial
(a)
C‐SPECT_WEHR10 10.0 10.1 6.8 10.1 6.8 10.5
C‐SPECT_WEHR15 10.2 10.3 6.9 10.2 6.9 10.5
A‐SPECT_LEHR 10.1 10.0 7.6 10.1 7.7 10.0
C‐SPECT_WEHR10 C‐SPECT_WEHR15 A‐SPECT_LEHR
FWHM (Average) 9.1 9.2 9.3
ASR C‐SPECT_WEHR10 C‐SPECT_WEHR15 A‐SPECT_LEHR
Center 0.99 0.99 1.01
X‐direction 1.49 1.48 1.33
Y‐direction 1.54 1.52 1.30
FWHM Center x‐direction y‐direction
Tangential Radial Tangential Radial Tangential Radial
(b)
C‐SPECT_WEHR 10.5 10.6 7.7 10.9 7.2 11.0
C‐SPECT_MEHRS 14.6 15.2 9.8 14.8 9.9 15.5
A‐SPECT_LMEGP 14.8 14.7 10.0 14.7 10.2 15.0
C‐SPECT_WEHR C‐SPECT_MEHRS A‐SPECT_LMEGP
FWHM (Average) 9.7 13.3 13.2
Aspect ratio C‐SPECT_WEHR C‐SPECT_MEHRS A‐SPECT_LMEGP
Center 0.99 0.96 1.01
x‐direction 1.42 1.51 1.48
y‐direction 1.59 1.57 1.46