Skip to main content
. 2021 Jan 10;22(2):165–177. doi: 10.1002/acm2.13158

Table 5.

Results of concentration linearity in each system. (a) 99mTc formulation; (b) 123I formulation.

Slope Intercept R2 Scatter fraction (MBq/mL)
(a)
C‐SPECT_WEHR10 (99mTc_SC7) 1024 6.030 0.9994 0.0059
C‐SPECT_WEHR10 (99mTc_SC3) 1044 7.877 0.9984 0.0075
C‐SPECT_WEHR15 (99mTc_SC7) 1156 7.541 0.9991 0.0065
C‐SPECT_WEHR15 (99mTc_SC3) 1219 9.429 0.9995 0.0077
A‐SPECT_LEHR 1117 11.61 0.9995 0.0104
Slope Intercept R2 Scatter fraction (MBq/mL)
(b)
C‐SPECT_WEHR (123I_SC7) 1433 4.897 0.9922 0.0034
C‐SPECT_WEHR (123I_SC3) 1579 6.642 0.9914 0.0042
C‐SPECT_MEHRS (123I_SC7) 1585 1.199 0.9952 0.0008
C‐SPECT_MEHRS (123I_SC3) 1656 1.437 0.9975 0.0009
A‐SPECT_LMEGP 1995 4.917 0.9995 0.0025