Table 3. Inter- and Intra-batch Reproducibility of Full Process Replicates for LC-MS Quantification of Estrogen Metabolites.

On each of 3 days, three replicate samples of charcoal stripped human urine (CSHU) were spiked with 250, 1000 and 6400 pg/mL estrogen metabolites. Variability in LC-MS measurements are reported as CV (%). CV values should not exceed 20%; if it does then the source of the variation should be identified [30].

Metabolite Inter-Batch CV (%) Day 1 Intra-Batch CV (%) Day 2 Intra-Batch CV (%) Day 3 Intra-Batch CV (%)

250 pg/mL 1000 pg/mL 6400 pg/mL 250 pg/mL 1000 pg/mL 6400 pg/mL 250 pg/mL 1000 pg/mL 6400 pg/mL 250 pg/mL 1000 pg/mL 6400 pg/mL
E3 13 10.1 12.5 1.4 1.2 3.5 5.8 7.2 3 4.2 2.2 7.6
E1 9.5 7.5 9.7 0.4 3.1 3.9 4.2 4.6 4.1 7.6 3 7.3
2-OHE1 7.1 5.5 6.9 3.1 4.6 3.3 3.6 7.7 4.9 4.9 0.9 7.8
16-ketoE2 5.3 5 7.3 0.7 1 4.8 6.6 6.5 3.4 5.7 6.2 6.2
4-OHE1 4.4 4.5 7.1 3.1 3 3.9 3.5 6.4 5 5.6 3.2 8
16α-OHE1 11 8.1 10.6 1.3 1.7 4.3 4.9 5.3 4.7 5.3 2.3 7
2-MeOE1 15.9 5.7 6.8 16.1 8.5 5.2 4.6 6.3 3.1 5.3 1.6 7.1
4-MeOE2 5.6 3.5 6.8 1.2 2.3 4.7 4.3 4.7 2.2 9.3 2.1 7.7
E2 13.6 10.5 10.9 1.7 1.7 4.5 6.4 6.8 3.6 4.6 2.5 8.4
16-epiE3 5.9 16.1 15.4 2.3 6 3.9 4.6 11.5 7.8 4.4 4.7 7.9
17-epiE3 16.1 20.1 15.6 4.8 6.4 6.3 6.8 12 8.5 14.9 20.1 4.5
4-MeOE1 12.7 11.5 11.4 0.7 1.9 4.5 4.3 4.4 2.8 5.1 2.7 5.8
2-OHE2 11.9 10.7 9.9 2.4 2.1 3.7 6.2 7 3 7.7 6.8 8.8
3-MeOE1 10.5 8.6 9.3 0.9 6.1 3.8 5 6 2.9 4 5.1 7.3
2-MeOE2 9.8 7.4 8.4 1 2.8 3.7 5.7 5.2 3.3 5.7 1.2 5.6
*

No standard was available for 17-epiE3, used D2-16-epiE3.