TABLE 3.

Reliability and reproducibility of the ergospirometry data obtained for the 40 participants performing cardiopulmonary exercise test with and without CLE test set-up

Parameter Agreement Lower LoA Upper LoA ICC (95% CI) WS-sd CoV %
VO2peak mL·min−1 11.4 −263 286 0.988 (0.977, 0.993) 98 2.6
VO2peak mL·kg−1·min−1 0.2 −3.5 4.0 0.957 (0.920, 0.977) 1.3 2.5
VCO2peak mL·min−1 53.1 −431 537 0.970 (0.943, 0.984) 176 3.8
VT L 0.0 −0.47 0.46 0.933 (0.876, 0.964) 0.16 6.3
VE L·min−1 2.6 −21.4 26.5 0.933 (0.877, 0.964) 8.7 6.2
HR beats·min−1 1.4 −11.6 14.3 0.720 (0.532, 0.841) 4.7 2.5
RR breaths·min−1 −2.5 −19.9 15.0 0.614 (0.375, 0.777) 6.0 11.1
RER units 0.01 −0.09 0.11 0.502 (0.233, 0.700) 0.04 3.1

CLE: continuous laryngoscopy during exercise; LoA: limit of agreement; ICC: intraclass correlation coefficient; WS-sd: within-subject sd; CoV: coefficient of variation; VO2peak: oxygen consumption at peak exercise; VCO2peak: carbon dioxide production at peak exercise; VT: tidal volume; VE: minute ventilation; HR: heart rate; RR: respiratory rate; RER: respiratory exchange rate.