Skip to main content
. 2021 Feb 15;8:14. doi: 10.1186/s40658-021-00361-y

Table 1.

Standard clinical reconstruction parameters for the different PET scanners

Iterative reconstruction Scanner type Iterations × subsets Gauss filter (mm) Voxel size (mm3)
OSEM Vision 10 × 5 4 1.7 × 1.7 × 2.0
mCT 3 × 24 3 2.0 × 2.0 × 2.0
mMR 3 × 21 4 2.1 × 2.1 × 2.0
OSEM-TOF Vision 4 × 5 4 1.7 × 1.7 × 2.0
mCT 2 × 21 3 2.0 × 2.0 × 2.0
mMR
OSEM-TOF+PSF Vision 4 × 5 4 1.7 × 1.7 × 2.0
mCT 2 × 21 3 2.0 × 2.0 × 2.0
mMR