Table 3.
Intra-day and Inter-day accuracy and precisions of GA and GABA calibration standards.
| Combined Intra-Day and Inter-Day Accuracy and Precision | |||||||||
| GA |
GABA |
||||||||
| Conc. | Measured Conc. | SD | % Error | % CV | Conc. | Measured Conc. | SD | % Error | % CV |
| 0.5 | 0.5 | 0.02 | −3.3 | 3.2 | 0.5 | 0.5 | 0.02 | 0.0 | 3.5 |
| 2.5 | 2.5 | 0.18 | −1.6 | 7.4 | 2.5 | 2.5 | 0.03 | 1.2 | 1.2 |
| 5 | 4.8 | 0.22 | −3.3 | 4.6 | 5 | 4.9 | 0.03 | −1.8 | 0.7 |
| 10 | 10.8 | 0.21 | 7.8 | 2.0 | 10 | 10.4 | 0.33 | 4.2 | 3.1 |
| 25 | 25.9 | 1.07 | 3.9 | 4.1 | 25 | 23.7 | 0.20 | −5.2 | 0.8 |
| 50 | 47.4 | 1.51 | −5.2 | 3.2 | 50 | 51.3 | 0.36 | 2.7 | 0.7 |
| 100 | 101.1 | 0.59 | 1.1 | 0.6 | 100 | 97.9 | 2.69 | −2.1 | 2.7 |
| Instrument Precision and Derivatization Reproducibility | |||||||||
| GA |
GABA |
||||||||
| Conc. | Measured Conc. | SD | % Error | % CV | Conc. | Measured Conc. | SD | % Error | % CV |
| 25 | 25.4 | 1.59 | 1.4 | 6.3 | 25 | 24.1 | 0.3 | −3.7 | 1.4 |