Published in final edited form as: Trends Analyt Chem. 2020 Oct 20;132:116049. doi: 10.1016/j.trac.2020.116049

Table 1.

Electrochemical techniques for tonic dopamine measurement.

Technique Temporal resolution Limit of detection (nM) Reported [DA]tonic Hardware & software Ref.
Chronoamperometry 1 min 19 ± 6 86 ± 6 nM, rat striatum E-chempro [115]
FSCAV 20 s 3.7 ± 0.5 90 ± 9 nM, mouse NAc NI DAQ LabVIEW [38]
CBM-FSCV 10 s 5.8 ± 0.9 >73 ± 5 nM,a rat striatum NI DAQ, LabVIEW, Matlab [39]
Convolution-based current removal 1 s <40a 41 ± 13 nM, rat NAc NI DAQ, LabVIEW [133]
M-CSWV 10 s 0.17 ± 0.03 120 ± 18 nM, rat striatum NI DAQ, WINCS, LabVIEW [42]
SWV 15 s 2.03 ± 0.09 82 ± 6 nM, rat striatum N/Aa, Matlab [43]
a

not reported in original manuscript.