Skip to main content
. 2021 Feb 16;136:104764. doi: 10.1016/j.jcv.2021.104764

Table 2.

Sensitivity, specificity and predictive values of RT-LAMP with raw sample input, compared to standard RT-PCR diagnosis.

RT-LAMP
RT-PCR Positive Negative N
Positive 35 (70%) 15 (30%) 50
< 20 cycles 24 (100%) 0 24
20−30 cycles 11 (65%) 6 (35%) 17
> 30 cycles 0 9 (100%) 9
Negative 0 51 (100%) 51
Result (95% CI)
Sensitivity 77% (65−86%)
Specificity 100% (90−100%)
NPV 70% (60−78%)
PPV 100 %
Accuracy 85% (76−91%)