Table 1.

Sensitivities and specificities of ELISA, LFIA and CLiA assays for SARS-CoV-2 antibody detection shown with 95% confidence intervals (CI)

  Number of Serum Samples
   
  RT-PCR confirmed cases
Pre-COVID control
   
ASSAY True Positive False Negative True Negative False Positive Sensitivity (95% CI) Specificity (95% CI)
ELISA            
EDITM 67 14 49 1 82.7(72.4–89.9) 98.0 (88.0–99.9)
DIA.PRO 83 7 42 8 92.2(84.1–96.5) 84.0 (70.3–92.4)
NOVALisa 67 22 50 0 75.3(64.8–83.5) 100.0 (91.1–100)
Euroimmun NC 81 9 50 0 90.0(81.4–95.0) 100.0 (91.1–100)
Euroimmun S1 80 10 50 0 88.9(80.1–94.3) 100.0 (91.1–100)
LFIA            
CTK Biotech 62 16 20 0 79.5(68.5–87.5) 100.0 (80.0–100)
CLIA            
Roche 68 7 25 0 90.7(81.1–95.8) 100.0 (83.4–100)

NC = Nucleocapsid. S1 = Spike Glycolipid 1