Table 5.

Statistical parameters of the measured cytokine concentrations, %CV between replicate runs, and the FEAD (BLQ%) across the technology platforms for serum samples derived from the healthy controls*.

Simoa (Quanterix) MESO V-Plex (Mesoscale Discovery) Luminex xMAP (Myriad) Luminex
(R&D Systems)
Quantikine
(R&D Systems)
Cytokine Measure HCs HCs HCs HCs HCs
IL-1ß Median (St Dev) 0.5 (0.05) 0.81 (1.51) 3.23 (0.47) 0.23 (0.47) 0.07 (0.02)
Min-Max 0.04 – 0.07 33.2–136.2 2.9–3.9 0–6.15 0.02–0.095
%CV (St Dev) 25 (10.8) 85.1 (50.0) NA 0.0 (2.9) 20.6 (15.5)
BLQ % 40% 80% 100% 100% 100%
IL-6 Median (St Dev) 1.59 (0.80) 1.0 (0.41) 2.30 (0.0) 2.63 (0.70) 3.03 (4.60)
Min-Max 0.92–3.05 0.53–1.65 2.3 1.27–3.20 1.36 (3.58)
%CV (St Dev) 0.3 (0.1) 4.5 (4.0) NA 0.0 (2.1) 1.0 (2.1)
BLQ % 0% 100% 100% 20% 0%
IFN-γ Median (St Dev) 0.18 (0.11) 2.76 (0.66) 2.0 (0.0) NA 3.2 (2.10)
Min-Max 0.07–0.35 3.55 (0.85) 2 NA 0.5–5.6
%CV (St Dev) 7.3 (11.2) 12.4 (39.0) NA 2.7 (3.3) NA
BLQ % 80% 100% 100% 100% 100%
TNF-α Median (St Dev) 0.95 (0.60) 1.07 (0.69) 21.33 (10.66) 5.94 (3.33) 0.35 (0.32)
Min-Max 0.91–2.3 0.54–2.6 11.0–36.0 0.8–10.62 0.20–1.10
%CV (St Dev) 13.9 (3.7) 2.0 (3.3) NA 1.9 (1.6) 2.1 (6.4)
BLQ % 0% 80% 100% 20% 0%

High: Precision, CV ≤ 20%; FEAD (BLQ%) < 50%.

Intermediate: Precision, CV 20–75%; FEAD (BLQ%) 50–75%.

Low: Precision, CV > 75%; FEAD (BLQ%) > 75.

*

Assay performance was ranked as follows.