Skip to main content
. 2020 Apr 28;2(2):100027. doi: 10.1016/j.cytox.2020.100027

Table A2.

Intra-assay precision (%CV) across the technology platforms for paired technical replicates of healthy control samples plated at different positions within a plate.

Cytokine Replicate Type Simoa (Quanterix) MESO V-Plex (Mesoscale Discovery) Luminex xMAP (Myriad) Luminex (R&D Systems) Quantikine (R&D Systems)
IL-1beta Plasma Pool 1 10.5 NA NA 51.2 23.4
Plasma Pool 3 3.9 111.0 NA 8.5 57.6
Healthy Control 5 4.8 95.7 NA 0.0 9.9
IL-6 Plasma Pool 1 4.6 2.5 NA 7.1 0.2
Plasma Pool 3 3.8 20.3 NA 5.3 0.9
Healthy Control 5 4.6 12.4 NA 4.9 11.0
IFN-γ Plasma Pool 1 9.5 NA NA NA
Plasma Pool 3 1.2 57.9 33.3 NA NA
Healthy Control 5 16.2 3.3 NA NA NA
TNF-α Plasma Pool 1 7.4 9.0 NA 5.4 2.1
Plasma Pool 3 2.8 1.9 43.3 3.2 3.6
Healthy Control 5 14.2 2.7 0.0 1.5 1.4