Skip to main content
. 2020 Apr 28;2(2):100027. doi: 10.1016/j.cytox.2020.100027

Table A3.

Inter-assay precision (%CV) across the technology platforms for paired technical replicates of healthy control samples on plate 1 vs plate 2.

Cytokine Replicate Type Simoa (Quanterix) MESO V-Plex (Mesoscale Discovery) Luminex xMAP (Myriad) Luminex (R&D Systems) Quantikine (R&D Systems)
IL-1beta Plasma Pool 1 0.1 129.3 NA 23.6 18.1
Plasma Pool 2 10.7 NA NA 19.3 35.9
Serum Pool 1 23.0 NA NA 12.5 BLD
IL-6 Plasma Pool 1 1.2 8.5 NA 3.9 1.0
Plasma Pool 2 7.0 1.8 NA 9.0 1.0
Serum Pool 1 10.2 1.6 NA 34.5 1.9
IFN-γ Plasma Pool 1 2.1 83.7 NA NA 14.3
Plasma Pool 2 7.5 53.4 NA NA 75.0
Serum Pool 1 13.3 16.4 0.0 NA NA
TNF-α Plasma Pool 1 0.1 6.2 NA 2.7 2.8
Plasma Pool 2 7.5 0.8 NA 8.4 0.4
Serum Pool 1 23.0 12.2 61.3 2.5 5.2

*Assay Performance was ranked as follows:

High: Precision, CV < 20%.

Intermediate: Precision, CV 20–75%.

Low: Precision, CV > 75%.