Skip to main content
. 2020 Dec 29;8(2):374–384. doi: 10.1002/acn3.51280

Table 2.

ROC analysis of individual RT‐QuICR assay parameters for PD diagnosis.

Sensitivity Specificity PPV NPV AUC (ROC) Threshold
Max Fluor 97.2% (92.1‐99.4) 94.1% (86.8‐98.1) 99.5% (89.7‐98.5) 96.4% (89.8‐99.2) 94.28% >78462 au
AUC‐fluor 96.3% (90.8‐99.0) 94.1% (86.8‐98.1) 95.4% (89.6‐98.5) 95.2% (88.3‐98.7) 94.87% >1441404 au⋅h
T50 93.5% (87.1‐97.4) 78.8% (68.6‐86.9) 84.9% (77.2‐90.8) 90.5% (81.5‐96.1) 88.99% <34.30 h
Lag 98.1% (93.5‐99.8) 83.5% (73.9‐90.7) 88.3% (81.2‐93.5) 97.3% (90.5‐99.7) 91.55% <27.85 h

Ranges in parentheses are 95% confidence intervals.

Abbreviations: au, arbitrary units; AUC, area under curve; AUC‐fluor, AUC of fluorescence versus time plot; Lag, reaction time to threshold; Max Fluor, maximum fluorescence; NPV, negative predictive value; PPV, positive predictive value; ROC, receiver operator characteristics; T50, reaction time to reach half‐maximal fluorescence.