. 2020 Oct 15;35(3):1453–1464. doi: 10.1007/s00464-020-08043-z

Table 2.

Damage score shows that LigaSure™ FC has less of an effect than electrosurgery’s SC mode

Device Power Energy device average activation time
1 s (n = 3) 3 s (n = 3) 5 s (n = 3)
50 W 2 2 2
Electrosurgery (Soft coagulation) 60 W 1.33 ± 0.47 2 2
70 W 2 2.33 ± 0.47 2
LigaSure (Flat coagulation) 0 1.33 ± 0.47 1.67 ± 0.47

LigaSure™ FC has a gradually increasing damage score after 3 s, but it is lower than that of electrosurgery’s SC mode (damage score at 3 s, LigaSure™ FC vs electrosurgery’s SC mode 60 W, 1.33 vs 2.0; p = 0.184)