Table 2. . Comparison of end point dPCR results with real-time dPCR results on 12 different negative clinical samples.

Sample FAM+ (real-time dPCR) FAM+ (end point dPCR)
S1 0 2
S2 0 2
S3 0 1
S4 0 2
S5 0 1
S6 0 2
S7 0 0
S8 0 0
S9 0 2
S10 0 2
S11 0 2
S12 0 8
Mean 0 2.00
SD 0 1.96
CV NA 97.89%