Areas under the receiver-operating characteristic curves (AUCs) and cutoff values for assessing pain in the spectrogram-convolutional neural network index and surgical pleth index (SPI) in surgical patients.

Parameter Spectrogram–CNNa index SPI
Training set Validation set Test set
AUC (95% CI) 0.992 (0.991-0.993) 0.932 (0.921-0.942) 0.757 (0.746-0.768) 0.659 (0.646-0.671)
P value <.01 <.01 <.01 <.01
Cutoff valueb 50 54 48 44
Sensitivity, %, specificity, % 94.6, 96.4 83.1, 88.2 68.1, 73.8 65.2, 59.5
Balanced accuracy, %c 95.5 85.7 71.0 62.4

aCNN: convolutional neural network.

bCutoff values used for the calculation of sensitivity and specificity were calculated as “best fit” (highest combined sensitivity and specificity).

cBalanced accuracy is the corrected accuracy of the imbalance of a class set, calculated as (sensitivity + specificity)/2.