. 2020 Mar 6;26(2):346–353. doi: 10.1111/hae.13949

Table 2.

Geometric mean (GM) and intra‐assay variability expressed as geometric coefficient of variation (%GCV) for all reagents/kits when testing plasma‐derived FIX (PD), recombinant FIX (RB) and dalcinonacog alfa (DA) against the plasma‐derived concentrate international standard: S1, using different one‐stage clotting assays (OSCA) reagents and chromogenic assay (ChA) kits. No %GCV is given for APTT‐SP as only 1 valid assay was obtained using this reagent

  PD RB DA
GM IU/mL GCV (%) GM IU/mL GCV (%) GM IU/mL GCV (%)
OSCA reagents
APTT‐SP 9.7 3.9 9.2 2.9 8.2
SynthasIL 9.7 5.7 9.4 13.6 8.9 6.8
SynthaFax 9.7 3.7 6.6 8.2 5.2 13.3
CK Prest 9.6 4.2 9.4 5.7 14.1 4.5
Cephascreen 9.5 6.6 10.6 6.2 11.1 3.0
Dapttin 9.9 8.4 9.7 6.8 13.8 10.5
Siron 10.3 4.8 10.1 9.9 9.4 8.7
Actin‐FS 10.1 5.9 9.7 4.1 7.6 4.5
Pathromtin SL 10.3 7.3 8.4 13.8 7.1 9.5
ChA kits
Rossix 10.3 2.1 7.6 4.0
Hyphen 9.5 4.0 6.6 6.9