TABLE 2.

Reproducibility and Reliability Based on the Manual Segmented SN Volumes of the NM‐MRI Sequences

GRE MT on TSE MT on TSE MT off
Mean volume mm3 (SD)
Rater 1 (test) 355.52 (54.53) 337.93 (64.73) 328.15 (69.43)
Rater 1 (retest) 373.91 (56.48) 362.99 (67.62) 306.11 (48.99)
Rater 2 (test) 333.87 (35.48) 343.10 (55.18) 302.84 (42.80)
Rater 2 (retest) 308.35 (36.73) 340.14 (45.73) 304.31 (37.37)
Rater 3 (test) 365.56 (52.83) 386.50 (42.14) 378.63 (36.38)
Rater 3 (retest) 391.11 (40.90) 414.67 (40.96) 380.67 (47.00)
Raters combined (test) 351.65 (48.73) 355.85 (57.49) 336.54 (59.30)
Raters combined (retest) 357.79 (57.03) 372.50 (60.08) 330.36 (56.41)
Test–retest variability (SD)
Rater 1 14.31% (13.08) 10.74% (10.16) 13.03% (12.76)
Rater 2 13.04% (10.68) 10.99% (7.84) 17.88% (9.82)
Rater 3 14.47% (12.05) 11.47% (9.84) 12.76% (7.83)
Test–retest ICC (95% CI)
Rater 1 0.13 (–0.48–0.66) 0.73 (0.26–0.92) 0.43 (–0.19–0.81)
Rater 2 0.05 (–0.54–0.61) 0.57 (–0.11–0.86) –0.13 (–0.66–0.48)
Rater 3 –0.10 (–0.64–0.51) 0.11 (–0.50–0.65) –0.01 (–0.59–0.57)
Rater ICC (95% CI)
Intrarater ICC (R1) 0.84 (0.67–0.93) 0.87 (0.71–0.94) 0.64 (0.32–0.83)
Intrarater ICC (R2) 0.33 (–0.9–0.65) 0.35 (–0.9–0.67) 0.53 (0.17–0.77)
Intrarater ICC (R3) 0.82 (0.30–0.94) 0.72 (0.45–0.87) 0.58 (0.12–0.81)
Interrater ICC 0.11 (–0.07–0.36) 0.34 (0.07–0.61) 0.21 (–0.01–0.47)

Given are the test–retest variability for each rater and ICC values with 95% confidence interval.

R1: rater 1; R2: rater 2; R3: rater 3; SD: standard deviation; ICC: intraclass correlation coefficient; 95% CI: 95% confidence interval; GRE: gradient recalled echo; MT: magnetization transfer pulse; TSE: turbo spin echo.