Table 9.
Position | Total signal [DN]a | Integration time [ms] | Thermal signal [DN]b | Rejectionc |
---|---|---|---|---|
Az: 10∘, El: 0∘ | 200 | 50 | .6 | 1.8e-5 |
Az: 20∘, El: 0∘ | 15 | 50 | .6 | 1.2e-6 |
Az: 30∘, El: 0∘ | 0.7d | 50 | .6 | <6e-8e |
Az: − 20∘, El: 0∘ | 200 | 50 | .6 | 1.8e-5 |
Az: − 20∘, El: 20∘ | 1450 | 10000 | 124 | 6e-8 |
aUnless otherwise noted, the measured signal is approximated as the respective Mastcam-Z’s estimate maximum value of the bias corrected frame
bThermal signal is approximated from the dark current reported in Sect. 3.3.2
cRejection is calculated as the ratio of the measured signal in [DN/ms] to the measured signal of the collimator: 226500 DN/ms
dAs no structure was observed in the frame to select a maximum, the frame average is used
eAs the signal is consistent with the thermal noise, we place a limit of the rejection percentage to be less than that generated by thermal noise alone