Fig. 7.

Fig. 7

Validation of the spatial resolution enhancement [36]. (A) Lateral resolution (Rlateral) measurements of OR-PAM and GR-PAM. The edge spread functions (ESFs) were obtained based on the PA amplitude across the sharp edge and the derivatives of ESFs give rise to the line spread functions (LSFs). The FWHM of ESFs for GR-PAM is smaller than OR-PAM by a factor of 1.6, which agrees with the theoretical value of2. (B) Axial resolution was estimated for both GR-PAM and OR-PAM. Raxial_GR was assessed based on the FWHM of the fitted Gaussian curve of the differential PA signal (PAGR-PAM) at different axial positions of a monolayer of RBCs. Raxial_OR was obtained from the FWHM of the A-line signal envelope (dashed line). Raxial_GR is measured to be 2.3 μm, which close to the theoretical prediction 2.4 μm (1.8λ/NA2) and more than 18 times finer than Raxial_OR. Figures reproduced with publisher’s permission [36].