Skip to main content
. 2020 Dec 8;85(5):2761–2770. doi: 10.1002/mrm.28629

TABLE 1.

Scan parameters of the two MP2RAGE sequences, which yielded the T1 maps, and the sequence parameters of the dynamic perfusion sequence

Precontrast MP2RAGE Postcontrast MP2RAGE Dynamic perfusion scan
TR/TE (ms) 5000/2.47 4000/2.32 3.7/1.3
TI1/TI2 (ms) 900/2750 900/2200 N/A
α12 (degrees) 5/3 4/5 6.5
Voxel size (mm3) 0.7 × 0.7 × 0.7 1.2 × 1.2 × 1.2 2.0 × 2.0 × 2.0
FOV (mm3) 168 × 224 × 224 192 × 230 × 230 192 × 192 × 32
Total acquisition time (minutes) 8:00 4:16 2:47
GRAPPA factor 3 3 3
Partial Fourier 6/8 6/8 6/8
Number of volumes 1 1 90
Time interval (seconds) N/A N/A 1.86