Skip to main content
. 2020 Dec 8;48(1):505–512. doi: 10.1002/mp.14605

Table I.

Range and Bragg peak width of a 148.2 MeV proton beam in water, determined using the in‐house developed water phantom (MR phantom) with and without magnetic fields, a commercial water phantom (reference), and Monte Carlo (MC) simulations.

Energy (MeV) Measurement Field strength (T) R80 (mm) BPW80 (mm)
148.2 MR phantom 0 150.2 4.1
1 149.4 4.1
Monte Carlo 0 150.2 4.2
1 149.5 4.2
Reference 0 150.2 4.2