Table 3:

Correlations between different performance metrics for all source control devices. R indicates the correlation coefficient while r2 is the coefficient of determination.

Correlation parameter #1 Correlation parameter #2 # points r r2 p-value
Cough aerosol collection efficiency Exhaled aerosol collection efficiency 22 0.8453 71% <0.0001
Manikin fit before coughing 22 0.5911 35% 0.0019
Filtration efficiency of material 22 0.7428 55% <0.0001
Airflow resistance of material 22 0.4280 18% 0.0235
Human subject fit (N95 mode) 22 0.7120 51% 0.0001
Human subject fit (all particle sizes) 22 0.8054 65% <0.0001
Exhaled aerosol collection efficiency Manikin fit before breathing 22 0.8863 79% <0.0001
Filtration efficiency of material 22 0.7199 52% 0.0001
Airflow resistance of material 22 0.4140 17% 0.0277
Human subject fit (N95 mode) 22 0.7528 57% <0.0001
Human subject fit (all particle sizes) 22 0.8585 74% <0.0001
Manikin fit before coughing Manikin fit before breathing 22 0.6389 41% 0.0007
Filtration efficiency of material Airflow resistance of material 22 0.6350 40% 0.0007