Table 3:
Correlations between different performance metrics for all source control devices. R indicates the correlation coefficient while r2 is the coefficient of determination.
Correlation parameter #1 | Correlation parameter #2 | # points | r | r2 | p-value |
---|---|---|---|---|---|
Cough aerosol collection efficiency | Exhaled aerosol collection efficiency | 22 | 0.8453 | 71% | <0.0001 |
Manikin fit before coughing | 22 | 0.5911 | 35% | 0.0019 | |
Filtration efficiency of material | 22 | 0.7428 | 55% | <0.0001 | |
Airflow resistance of material | 22 | 0.4280 | 18% | 0.0235 | |
Human subject fit (N95 mode) | 22 | 0.7120 | 51% | 0.0001 | |
Human subject fit (all particle sizes) | 22 | 0.8054 | 65% | <0.0001 | |
Exhaled aerosol collection efficiency | Manikin fit before breathing | 22 | 0.8863 | 79% | <0.0001 |
Filtration efficiency of material | 22 | 0.7199 | 52% | 0.0001 | |
Airflow resistance of material | 22 | 0.4140 | 17% | 0.0277 | |
Human subject fit (N95 mode) | 22 | 0.7528 | 57% | <0.0001 | |
Human subject fit (all particle sizes) | 22 | 0.8585 | 74% | <0.0001 | |
Manikin fit before coughing | Manikin fit before breathing | 22 | 0.6389 | 41% | 0.0007 |
Filtration efficiency of material | Airflow resistance of material | 22 | 0.6350 | 40% | 0.0007 |