Skip to main content
. 2021 Feb 19;27:1076029620976913. doi: 10.1177/1076029620976913

Table 2.

Concentrations of FVIII in Jivi®, as Determined by Clotting (With Various APTT Reagents) and Chromogenic Substrate Methods.

Jivi® X Methods A B C D E F G H I J K L
FVIII(IU) 1.00 1.01 APTT 0.22 1.06 0.96 1.01 0.83 0.41 1.08 0.85 0.91 1.10 0.82 0.08
1st DPT 0.10 1.30 1.10 1.20 0.56 0.54 1.10 0.98 1.10 1.30 1.02 0.08
1st DPH 0.04 1.88 1.20 1.46 1.24 1.34 1.30 1.26 1.16 1.34 1.70 0.13
0.50 0.43 APTT 0.09 0.54 0.50 0.51 0.46 0.19 0.54 0.44 0.44 0.50 0.38 0.05
1st DPT 0.04 0.62 0.56 0.56 0.28 0.26 0.54 0.51 0.56 0.07 0.51 0.04
1st DPH 0.02 0.92 0.62 0.72 0.64 0.64 0.70 0.62 0.56 0.70 0.88 0.04
0.10 0.11 APTT 0.02 0.11 0.11 0.10 0.11 0.08 0.11 0.10 0.10 0.10 0.10 0.02
1st DPT 0.01 0.13 0.11 0.11 0.07 0.09 0.12 0.10 0.13 0.12 0.11 0.01
1st DPH 0.003 0.18 0.13 0.14 0.18 0.14 0.17 0.16 0.16 0.14 0.16 0.1.2
0.05 0.04 APTT 0.01 0.05 0.05 0.05 0.05 0.04 0.07 0.05 0.05 0.04 0.05 0.01
1st DPT 0.002 0.06 0.05 0.05 0.04 0.04 0.05 0.05 0.06 0.06 0.05 0.004
1st DPH 0.001 0.08 0.06 0.07 0.08 0.08 0.08 0.08 0.07 0.07 0.07 0.06
0.01 0.007 APTT 0.003 0.01 0.01 0.01 0.01 0.01 0.02 0.01 0.01 0.01 0.01 error
1st DPT error 0.01 0.01 0.01 0.01 0.01 0.01 0.01 0.01 0.01 0.01 0.001
1st DPH error 0.01 0.01 0.01 0.02 0.01 0.02 0.02 0.02 0.02 0.02 0.001
0.005 0.002 APTT error 0.005 0.005 0.006 error error 0.007 0.005 0.005 0.005 0.005 error
1st DPT error 0.006 0.005 0.006 error 0.004 0.006 0.005 0.006 0.006 0.005 error
1st DPH error 0.006 0.006 0.007 error 0.008 0.007 0.007 0.008 0.007 0.008 error

X, chromogenic substrate method; A, 2ndDPT; B-L, FFT; 1stDPT, first derivative peak time; FVIII, coagulation factor FVIII. 1stDPH, first derivative peak height;

APTT, activated partial thromboplastin time; 2nd DPT, second derivative peak time; [APTT reagents] A, APTT-SP; B, APTT-SS; C, STA Cephascreen; D, Coagpia APTT-N; E, APTT PSL; F, C.K. Prest; G, APTT-SLA; H, TC APTT; I, APTT FS; J, APTT FSL; K, PTT ACT; L, STA PTT A.