Skip to main content
. Author manuscript; available in PMC: 2022 Jan 1.
Published in final edited form as: J Cardiovasc Electrophysiol. 2020 Nov 13;32(1):138–147. doi: 10.1111/jce.14798

Table 3:

Comparison of visual scores between TV-ICDs and pacemakers across three pulse sequences. SVS: summed visual score.

Pacemaker TV-ICD p-value
Standard Segmented Conspicuity 4.5 [4.2, 4.6] 3.1 [2.5, 3.6] <0.001
Artifact 4.4 [4.3, 4.6] 3.1 [2.7, 3.5] <0.001
SVS 8.7 [8.6, 9.2] 6.2 [5.2, 7.1] <0.001
Wideband Segmented Conspicuity 4.4 [4.0, 4.8] 4.6 [4.2, 4.8] 0.977
Artifact 4.5 [4.2, 4.8] 4.6 [4.2, 4.8] 0.960
SVS 8.9 [8.1, 9.5] 9.3 [8.4, 9.6] 0.993
Wideband Single-Shot Conspicuity 4.5 [4.0, 5.0] 4.4 [3.9, 5.0] 0.424
Artifact 4.6 [4.2, 5.0] 4.4 [4.2, 5.0] 0.622
SVS 9.0 [8.3, 10.0] 8.8 [8.1, 10.0] 0.505