. 2021 Jan 25;9(2):245. doi: 10.3390/microorganisms9020245

Table 2.

Diagnostic assessment of the three automated analyzers for SARS-CoV-2 IgG detection using RT-PCR as a reference test for the (A) overall sample, (B) symptomatic patient group, and (C) asymptomatic patient group.

RT-PCR Overall Percent Agreement Sensitivity Specificity Cohen’s Kappa Statistic
Positive Negative Total % (95% CI) % (95% CI) % (95% CI) k (95% CI)
(A)
Negative 11 121 132
Total 111 127 238
VIDAS®3 assay Positive 98 2 100 93.7 (89.9–96.2) 88.3 (81.0–93.0) 98.4 (94.5–99.6) 0.87 (0.83–0.92)
Negative 13 125 138
Total 111 127 238
CL-900i® assay Positive 100 6 106 92.9 (88.9–95.5) 90.1 (83.1–94.4) 95.3 (90.1–97.8) 0.86 (0.81–0.90)
Negative 11 121 132
Total 111 127 238
LIAISON®XL assay Positive 95 0 95 93.3 (89.4–95.8) 85.6 (77.9–90.9) 100 (97.1–100) 0.86 (0.82–0.91)
Negative 16 127 143
Total 111 127 238
(B)
VIDAS®3 assay Positive 54 2 56 97.8 (94.5–99.1) 96.4 (87.9–99.0) 98.4 (94.5–99.6) 0.95 (0.92–0.98)
Negative 2 125 127
Total 56 127 183
CL-900i® assay Positive 55 6 61 96.2 (92.3–98.1) 98.2 (90.6–99.7) 95.3 (90.1–97.8) 0.91 (0.87–0.95)
Negative 1 121 122
Total 56 127 183
LIAISON®XL assay Positive 51 0 51 97.3 (93.8–98.8) 91.1 (80.7–96.1) 100 (97.1–100) 0.93 (0.90–0.97)
Negative 5 127 132
Total 56 127 183
(C)
Negative 10 121 131
Total 51 127 178
VIDAS®3 assay Positive 40 2 42 92.7 (89.8–96.2) 78.4 (64.7–88.7) 98.4 (94.5–99.6) 0.81 (0.75–0.87)
Negative 11 125 136
Total 56 127 178
CL-900i® assay Positive 41 6 47 91.0 (85.9–94.4) 80.4 (67.5–89.0) 95.3 (90.1–97.8) 0.77 (0.71–0.84)
Negative 10 121 131
Total 51 127 178
LIAISON®XL assay Positive 40 0 40 93.8 (89.3–96.5) 78.4 (65.4–87.5) 100 (97.1–100) 0.84 (0.78–0.90)
Negative 11 121 131
Total 51 127 178