. 2021 Feb 23;15:775–782. doi: 10.2147/OPTH.S300871

Table 2.

Validation for Drive-by Photoscreening; the Validation Statistics Sensitivity, Specificity, Positive Predictive Value (PPV) and ABCD Sensitivity (Sens) and ABCD Specificity (Spec) are Given in the Right-Hand Columns. A 3 x 2 Matrix of Validation Variables includes Conventional A, B, C and D with Additional E (Inconclusive Exam True) and F (Inconclusive Exam Normal). Exam Failure Criteria (Exam Fail?) Include Three Subsets; AAPOS 2003 Refractive Only (Refract Only), Cases with Visual Acuity Defined Amblyopia and/or Constant Strabismus (Strab/Ambly) and the AAPOS 2003 Which Combines Cycloplegic Refraction Guideline Levels Plus Strabismus Plus Amblyopia. There are Three Levels of Instrument Referral Criteria for plusoptiX and 2WIN Shown in Table 1 Offering Sensitive (Sens), Baseline (Refer) and Specific (Spec) Screen Interpretations

Drive-By Photoscreen Exam+ Exam-
refer A B
pass C D
inconclusive E F i=E+F A/(A+C) D/(B+D) A/(A+B) A/(A+C+i) D/(B+D+i)
maker model Exam Fail? criteria A B C D E F sensitivity specificity PPV ABCD-sens ABCD-spec
plusoptiX S12 AAPOS 03 sens 43 4 6 21 12 2 88% 84% 91% 68% 54%
AAPOS 03 refer 41 1 7 25 12 2 85% 96% 98% 66% 63%
AAPOS 03 spec 34 0 14 26 12 2 71% 100% 100% 55% 65%
strab-ambly refer 17 24 3 30 6 8 85% 56% 41% 50% 44%
refract only refer 41 1 5 27 11 3 89% 96% 98% 68% 64%
Adaptica 2WIN AAPOS 03 sens 51 5 6 22 3 1 89% 81% 91% 84% 71%
Kaleidos AAPOS 03 refer 44 3 12 25 3 1 79% 89% 94% 73% 78%
AAPOS 03 spec 38 3 19 24 3 1 67% 89% 93% 62% 77%
strab-ambly refer 21 27 5 31 1 3 81% 53% 44% 70% 50%
refract only refer 44 4 10 26 3 1 81% 87% 92% 76% 76%
Rebion blinq AAPOS 03 refer 23 7 31 20 7 0 43% 74% 77% 38% 59%
strab-ambly refer 13 17 11 40 3 4 54% 70% 43% 42% 63%
refract only refer 21 9 29 2 7 0 42% 18% 70% 37% 11%