Huntington disease (HD) reference samples with published allele sizes [30] compared to those determined using the AmplideX PCR/CE HTT Kit.

Vendor ID Input (ng) Categorical Bin Published
(Kalman et al., 2007)
Measured
AmplideX PCR/CE HTT Kit
Allele 1 Allele 2 Allele 1 Allele 2
NA20245 20 Normal 15 15 15 15
NA20206 20 Normal 17 18 17 18
NA20207 20 Normal 19 21 19 21
NA20246 20 Normal 15 24 15 24
NA20247 20 Intermediate 15 29 15 29
NA20248 20 Reduced Penetrance 17 36 17 36
NA20249 20 Reduced Penetrance 22 39 22 39
NA20208 20 Expanded 35 45 35 45
NA20209 20 Expanded 45 47 45 47
NA20210 20 Expanded 17 74 17 75
NA20250 20 Expanded 15 40 15 40
NA20251 20 Expanded 39 50 39 50
NA20252 20 Expanded 22 65 & 66 22 66 *
NA20253 20 Expanded 22 100 22 100 **

Allele genotypes for Coriell samples shown from Kalman et al. [30] are the mode; error limits were set at ±1 for alleles ≤42 repeats and ±3 repeats for alleles >42 CAG repeats, according to the European Molecular Genetics Quality Network best practice guidelines for the molecular genetic testing of Huntington disease [31]. * A 63 CAG mosaic allele was also detected. ** A 128 CAG mosaic allele was also detected. Underlined and in bold, the difference between the two methodologies.