Comparison the data from the present study of the angle deviation and the global deviation at the coronal end of the implants between dynamic and static navigation and a work on freehand implant placement (mean value and standard error (if specified)).

Study Angle Deviation Global Coronal Deviation Global Apical Deviation
Mean Mean Mean
Dynamic navigation in vitro 4.1° 1.03 mm 1.04 mm
Dynamic navigation clinical 3.7° 1.00 mm 1.33 mm
Static navigation—clinical review 1 [65] 3.6° 1.10 mm 1.40 mm
Static navigation—clinical review 2 [10] 3.5° 1.2 mm 1.40 mm
Freehand implant placement [66] 9.9° 2.77 mm 2.91 mm