Skip to main content
. 2021 Feb 12;11(2):297. doi: 10.3390/diagnostics11020297

Table 4.

Clinical specimens with discrepancies between Xpert HBV Viral Load, CAP/CTM HBV version 2.0, Abbott RealTime HBV, and VERIS MDx HBV assay results.

Specimen ID Storage Xpert HBV VL
(Log10 IU/mL)
CAP/CTM HBV 2.0
(Log10 IU/mL)
Difference
(Log IU/mL)
Abbott RealTime HBV/VERIS MDx HBV
(Log10 IU/mL)
HBV364192 1.18 3.76 −2.58 1.18
HBV348238 1.38 2.68 −1.30 2.59
HBV229152 * 1.63 2.64 −1.01 3.10
HBV380197 1.86 3.27 −1.40 3.52
HBV364256 2.07 3.15 −1.08 2.21
HBV229150 * 2.82 1.62 1.20 2.52
HBV380249 3.09 1.36 1.73 2.98
HBV348195 4.30 3.05 1.25 4.13
HBV348255 4.62 5.88 −1.26 4.88
HBV348220 5.70 3.43 2.27 5.29
HBV364494 6.83 5.67 1.17 6.50
HBV364384 7.70 5.60 2.10 7.48
HBV348100 7.79 6.73 1.06 7.70

Note: * tested using the VERIS HBV assay; all others tested using the Abbott HBV assay.