Skip to main content
. 2021 Jan-Feb;36(1):48–56. doi: 10.21470/1678-9741-2020-0115

Table 2.

Procedure-related data.

  n % Literature***
Procedure time (min, mean±SD, range) 104±31(45-186)    
Echocardiography, n (%)      
Transesophageal 59/59 100%  
Success 57/59 96.6% 92%
    (95% CI: 0.90-0.94)
Device sizes, n (%)      
5 7/57 12%  
6 15/57 26%  
7 17/57 30%  
8 10/57 18%  
10 5/57 9%  
12 3/57 5%  
Converting to surgical repair 2/59 3.4% 80/1368
(5.8%)
Clinically insignificant residual shunt 2/59 3.4% 95/1368
(6.9%)
Total severe complications 3/59 5.3% 109/1368
-8%
Intraoperative complications 2/59 3.4% 88/1368
-6%
(95% CI: 0.028-0.071)
Significant residual shunt* 1/59 1.7% 32/1368
New-onset aortic valve insufficiency* 1/59 1.7% 31/1368
Complete heart block None   10/1368
Failure to establish a path None   8/1368
New-onset tricuspid valve insufficiency None   7/1368
Postoperative early complications 1/57 1.7% 12/1368
0.9%
Significant residual shunt None   0
New-onset aortic valve insufficiency None   0
Complete heart block None   5/1368
New-onset tricuspid valve insufficiency None   1/1368
New-onset tricuspid valve insufficiency None    
Occluder dislodgement** 1/57 1.7% 2/1368
Second operation** 1/57 1.7% 4/1368
Follow-up period No major complications 0% 9/1368

iRBBB=incomplete right bundle branch block.

*

The patient's device was extracted and surgical closure was performed.

**

Same patient.

***

Hong et al.[5].