Table 2.

Characteristics of the five kits used for the detection of SARS-CoV-2 genomic targets.

Parameters Kit-A Kit-B Kit-C Kit-D ° Kit-E
GENES Gene E x x x x
RdRP x x x
Gene n x x
orf1ab x
INSTRUMENTS ABI 7500 x x x
LC 480 x x
Montania x
BioRad CFX96Dx x x x x x
Rotorgene 6000 x
Q-Qiagen x
Atila PG 9600
ABI QuantStudio 5DX
AriaDx x
Mic x
KIT FEATURES LoD 100 copies/rxn 25 copies/rxn 3 copies/rxn 1 copie/rxn 50 copies/rxn
Precision CV ≤ 5% not declared CV ≤ 5% not declared CV ≤ 5%
Treshold Ct ≤40 positive
≤40 IC
≤30 PC ≤32 IC ≤34 IC ≤25 positive ≤40 positive
Number of total cycles 45 35 45 35 45
Time ~1 h 30 min ~1 h ~1 h 30 min 20 min ~1 h 30 min
RNA Volume 8 μL 10 μL 10 μL 5 μL 5 μL OM1
5 μL OM2
Final Volume of reaction 25 μL 25 μL 30 μL 25 μL 25
Volume of Internal
Control
10 μL in the starting sample 5 μL when starting with extraction; 0.2 μL when spiked in RNA 1 μL in 6 µL elution volume Not included Not included

° Ionebio (Kit-D; http://www.ionebio.com (accessed on 4 February 2021)) is an isothermal-based method which determines as positive all samples with a Cycle thershold (Ct) ≤ 25. This kit works as a loop-mediated isothermal amplification (LAMP) method; CV: coefficient of variation; IC: internal control; OM: mastermix; PC: positive control. Kit E: https://www.theglobalfund.org/media/9629/covid19_diagnosticproducts_list_en.pdf (accessed on 4 February 2021).