Table 1.

CT Chest protocols for quantitative lung imaging

Siemens Definition AS+ Siemens FORCE GE CT750 HD GE Revolution Philips iCT Canon Genesis Canon
PrimeSP
Intravenous contrast No No No No No No No
Detector configuration (number x mm) 128 × 0.6 192 × 0.6 64 × 0.625 128 × 0.625 128 × 0.625 80 × 0.5 80 × 0.5
SOV Large body Large body Large Large
Organ characteristic Thorax
Ensured by first selecting a Siemens default routine adult chest protocol
Thorax
Ensured by first selecting a Siemens default routine adult chest protocol
Rotation time (s) 0.5 0,025 0.5 0.5 0.5 0.275 0.35
Scanner Pitch 1.0 1.0 0.984 0.992 1.0 Standard Standard
kVp 120 (Automated kV control – off) 120 (Automated kV control – off) 120 120 120 120 120
Tube current modulation Care dose: on Care dose: on AutomA & SmartmA: On AutomA & SmartmA: On DoseRight On SureExposure On SureExposure On
Quality reference mAs – Software v. 2012B, VA48, and VA50 42 36
Quality reference mAs – (VB10 & VB20) 29 25
Max mA 800 700 700
mA SmartmA 30–500 20–700 20–700
Noise index 85 61 12.5 12.5
DRI 8
Slice thickness x interval (mm) 1 × 0.5 1 × 0.5 1 × 0.5 1 × 0.5 1 × 0.5 1 × 0.5 1 × 0.5
Reconstruction Kernel Q30f Qr40 Standard Standard B (Standard) Lung Standard Lung Standard
Iterative algorithms SAFIRE 5 ADMIRE 5 ASiR-V 60% ASiR-V 60% iDose 6 AIDR3De AIDR 3De
Reconstruction mode Helical Plus Helical Plus
Additional filters IQ Enhance OFF IQ Enhance OFF
Approximate dose (for a 30 cm scan, average-size adult) CTDIvol: 2.46 mGy
Effective dose: 1.65 mSv
CTDIvol: 2.21 mGy
Effective dose: 1.49 mSv
CTDIvol 4.87 mGy
Effective dose: 3.27 mSv
CTDIvol: 3.38 mGy
Effective dose: 2.27 mSv
CTDIvol: 3.5 mGy
Effective dose: 2.35 mSv
CTDIvol: 2.4 mGy
Effective dose: 1.61 mSv
CTDIvol: 2.4 mGy
Effective dose: 1.61 mSv

CTDIvol, volume computed tomography dose index; DRI, dose right index ; SOV, scan field of view.