Needle placement accuracy (in mm)—Summary table of 3D, lateral and depth deviations-All operators.

Parameter N Mean (SD) Median (Q1 ; Q3) Min ; max LSmean [95% upper limit] P value*
3D deviation (mm) 43 4.19 (2.03) 3.70 (2.70 ; 5.30) 0.80 ; 10.20 4.15 [0 ; 4.74] P = 0.0138
36 3.50 (1.25) 3.40 (2.60 ; 4.35) 0.80 ; 5.90 3.59 [0 ; 3.99] P ≤ .0001
Lateral deviation (mm) 43 3.28 (1.80) 3.10 (2.00 ; 4.10) 0.60 ; 8.90 3.26 [0 ; 3.83] P = 0.0005
36 2.82 (1.18) 2.75 (1.85 ; 3.65) 0.60 ; 5.00 2.88 [0 ; 3.32] P ≤ .0001
Depth deviation (mm) 43 1.27 (2.62) 1.5 (0.00 ; 2.70) -5.00 ; 7.20 1.22 [0 ; 2.04] P ≤ .0001
36 1.07 (2.06) 1.40 (0.10 ; 2.20) -3.80 ; 6.20 1.05 [0 ; 1.86] P ≤ .0001

Analyses based on all or last needle insertion measure.

*P value for unilateral testing distance H0: > 5 mm vsersu H1: ≤ 5 mm.

Least Square (LS) Mean and 95% Confidence Interval (CI) taking into account intra-animal and fiducial nested in animal variability structure.