Table 2.

Error propagation from uncertainty of 23Na relaxation times. In the left section the mean TSC and CIC in an ROI containing the lesion (Patient 1 in Fig. 5) is reported for different values of 23Na T1. In the central section mean TSC and CIC in the lesion are reported for different values of T2,s. in the rightmost section mean CIC and TSC of the tumor are reported for a range of T2,l. In the last row the mean uncertainty, defined as mean percentage variation relative to the central values, is reported. The relaxation times that we set in our method and the corresponding mean quantification results in the lesion are highlighted in bold. Due to the scarcity of literature references for 23Na relaxation times in breast cancer, an abundant range of relaxation properties (± 50%) was simulated. Notations are: CIC intracellular sodium concentration, TSC total sodium concentration.

Input Output Input Output Input Output
T1 (ms) Lesion TSC (mM) Lesion CIC (mM) T2,s (ms) Lesion TSC (mM) Lesion CIC (mM) T2,l (ms) Lesion TSC (mM) Lesion CIC (mM)
16 41.2 18.5 0.25 57.2 43.8 8 47.3 30.2
24 43.6 22.3 0.38 50.3 33.5 12 47.1 29.8
32 47.0 29.6 0.5 47.0 29.6 16 47.0 29.6
40 50.9 34.6 0.63 44.9 24.4 20 47.0 29.6
48 55.1 41.7 1 41.9 22.2 24 47.0 29.6
Mean uncertainty ± 50% ± 14% ± 39% ± 50% ± 15% ± 33% ± 50% ± 0.03% ± 1%