Skip to main content
. 2020 Dec 29;69(2):385–390. doi: 10.4103/ijo.IJO_2333_20

Table 1.

Overall Median logMAR visual acuity with range (minimum, maximum) for COMPlog, SmartOptometry, Peek Acuity and Reduced Snellen near vision chart

Participants Median age [Min, Max] Charts Median Visual Acuity [Min, Max] P
All (n = 68) 31 [20, 60] COMPlog 0.01 [0.00,1.09] 0.31
Peek Acuity 0.00 [0.00,1.30]
SmartOptometry 0.00 [0.00,1.00] 0.002*
Reduced Snellen near vision 0.00 [0.00,0.80]
Young (n = 44) 27 [20, 36] COMPlog 0.00 [0.00,1.09] 0.98
Peek Acuity 0.00 [0.00,1.30]
SmartOptometry 0.00 [0.00,1.00] 0.004*
Reduced Snellen near vision 0.00 [0.00,0.78]
Presbyopia (n = 24) 50 [37, 60] COMPlog 0.10 [0.00,0.83] 0.17
Peek Acuity 0.15 [0.00,1.30]
SmartOptometry 0.39 [0.00,0.60] 0.007
Reduced Snellen near vision 0.30 [0.00,0.80]
Myopia (n = 29) 29 [20, 56] COMPlog 0.36 [0.00,1.09] 0.90
Peek Acuity 0.00 [0.00,1.30]
SmartOptometry 0.30 [0.00,1.00] 0.74
Reduced Snellen near vision 0.00 [0.00,0.80]
Hyperopia (n = 16) 50 [22, 60] COMPlog 0.06 [0.00,0.58] 0.74
Peek Acuity 0.10 [0.00,0.60]
SmartOptometry 0.40 [0.00,0.60] 0.11
Reduced Snellen near vision 0.30 [0.00,0.78]
Emmetropia (n = 16) 29 [21, 47] COMPlog 0.00 [0.00,0.04] 0.71
Peek Acuity 0.00 [0.00,0.10]
SmartOptometry 0.00 [0.00,0.40] 0.04*
Reduced Snellen near vision 0.00 [0.00,0.18]

*indicates statistically significant P-values (P < 0.05)