Skip to main content
. 2021 Feb 1;54(Pt 1):263–279. doi: 10.1107/S1600576720016192

Table 1. Time-averaged neutron intensities at the sample and different parts of the scattering characterization system with a pyrolytic graphite single-crystal sample simulated with McStas and Geant4 .

Position McStas (Hz) Geant4 (Hz)
Sample incident 1.45 × 1010 1.45 × 1010
Sample transmission 1.16 × 1010 1.16 × 1010
2.7 meV analyser 1.50 × 109 1.50 × 109
5.0 meV analyser 1.20 × 109 1.19 × 109
5.0 meV detector 1.98 × 108 1.98 × 108
Analyser transmission 7.47 × 108 7.31 × 108