Skip to main content
. 2021 Mar 11;36(2):441–447. doi: 10.1038/s41433-021-01485-2

Table 2.

Sensitivity and specificity of undilated Smartscope vs. dilated Topcon.

Undilated Smartscope vs. dilated Topcon
Undilated Smartscope G1 Dilated Topcon-G1 Sensitivity and specificity (CI)
Glaucoma No glaucoma
SP Glaucoma 128 5 Sensitivity: 97.71% (93.4–99.5%)
SN Glaucoma 3 140 Specificity: 96.55% (92.1–98.9%)
Total 131 145
Undilated Smartscope G2 Topcon-G2 Sensitivity and specificity (CI)
Glaucoma No glaucoma
SP glaucoma 128 4 Sensitivity: 95.52% (90.5–98.3%)
SN glaucoma 6 138 Specificity: 97.18% (92.9–99.2%)
Total 134 142

SP screening positive, SN screening negative, G grader.