Skip to main content
. 2020 Nov 19;33(6):293–299. doi: 10.1159/000511443

Table 3.

Three days precision and accuracy of the HPLC method for quantification of CoQ10

Standard Cnom, µg/mL Interday variability, n = 15 Intraday variability, n = 9
mean±SD, µg/mL CV, % RE, % mean±SD, pg/mL CV, % RE, %
LLOQ 0.0618 0.062±0.002 2.5 −0.1 0.061±0.002 2.8 −1.2
QC1 0.1854 0.187±0.004 1.9 1.1 0.187±0.003 1.7 0.6
QC2 4.12 4.223±0.069 1.6 2.5 4.224±0.016 0.4 2.5
QC3 8.24 8.532±0.149 1.8 3.5 8.643±0.042 0.5 4.9

QC1, 3 times the LLOQ; QC2 medium; QC3, 75% of the upper calibration curve range; SD, standard deviation; COQ10, coenzyme Q10.