Skip to main content
. 2020 Dec 9;35(3):e23669. doi: 10.1002/jcla.23669

Table 1.

The precision and accuracy for paraquat and diquat in human plasma

Analyte

Concentration

(μg/ml)

Intra‐day Inter‐day Extraction recovery

Mean ± SD

(μg/ml)

Accuracy

RE%

Precision

RSD%

Mean ± SD

(μg/ml)

Accuracy

RE%

Precision

RSD%

Mean ± SD

(%)

RSD

%

Paraquat 0.05 0.052 ± 0.004 4.0 7.07 0.056 ± 0.009 12.0 16.49
0.1 0.098 ± 0.010 −2.0 9.76 0.101 ± 0.011 1.0 11.06 95.38 ± 7.01 7.35
1.5 1.560 ± 0.066 4.0 4.20 1.484 ± 0.088 −1.1 5.90
15 14.634 ± 0.454 −2.4 3.10 14.580 ± 0.961 −2.8 6.60 103.97 ± 0.89 0.85
Diquat 0.05 0.052 ± 0.006 4.0 12.39 0.050 ± 0.007 0.0 14.14
0.1 0.097 ± 0.011 −3.0 11.30 0.099 ± 0.012 −1.0 12.60 94.79 ± 3.57 3.76
1.5 1.590 ± 0.057 6.0 3.60 1.504 ± 0.095 0.3 6.28
15 14.868 ± 0.478 −0.9 3.22 14.928 ± 1.074 −0.5 7.19 98.40 ± 0.79 0.81