Table 2:

Measured entrance air kerma rates for a typical patient examination and for the largest FOV in fluoroscopic mode II preferred for clinical imaginga

System DRT Plane kV Focal Spot (mm) Measured EAKR (mGy/min) a,r Ratio
FD 20/20 No AP 68 ± 1 Small (0.4) 5.1 ± 0.2 1.03 ± 0.12
LAT 69 ± 1 Small (0.4) 6.1 ± 1.0 1.00 ± 0.04
FD 20/10 Before AP 68 ± 1 Small (0.4) 4.9 ± 1.0 1.07 ± 0.03
LAT 72 ± 4 Small (0.5) 6.3 ± 3.4 0.97 ± 0.004
After AP 68 ± 1 Small (0.4) 2.8 ± 0.4 0.96 ± 0.08
LAT 73 ± 2 Small (0.5) 4.4 ± 0.6 1.01 ± 0.1

Note:—AP indicates anteroposterior; DRT, dose-reduction technology.

a

For each system, the selected kilovolt and x-ray focal spot along with its nominal size (millimeter) for each plane are summarized. For the FD 20/10 system, these values are reported before and after the installation of the DRT. The x-ray beam filtration is 0.4 mm Cu and 1 mm of Al for all systems, platforms, and planes.