Table 3:

DSA programmed settings for the image-receptor (detector) entrance air kerma rate for a typical patient examination with the largest FOV on both platformsa

System DRT Acquisition Protocol kV Filtration Focal Spot (mm) Programmed RAKR (μGy/frame)
FD 20/20 No Standard 80 0.1 mm Cu +1 mm Al Large (AP/LAT: 0.7 mm) 4.0
FD 20/10 Before Standard 80 0.1 mm Cu +1 mm Al Large (AP: 0.7 mm; LAT: 0.8 mm) 4.0
After Quarter 75 0.1 mm Cu +1 mm Al Small (AP: 0.4 mm; LAT: 0.5 mm) 0.7
Half 78 No added filtration Small (AP: 0.4 mm; LAT: 0.5 mm) 1.0
Full 80 0.1 mm Cu +1 mm Al Large (AP: 0.7 mm; LAT: 0.8 mm) 4.0

Note:—AP indicates anteroposterior; DRT, dose-reduction technology.

a

The dose-reduction platform was equipped with 3 acquisition protocols in which the “full-dose” protocol reverts to the reference platform hardware and software settings (“standard” dose protocol). The programmed settings are identical for both planes in each system and for each acquisition protocol.