Table 1:

Angiographic results of clot retrieval

Vessel Animal Number
1
2
3
4
5
6
R APA L APA R APA L APA R APA L APA
Vessel diameter 2 2.5 2 2.1 2.1 2.3
Clot type Soft Soft Soft Firm Firm Firm
TIMI postdevice deploymenta 0 2 N/Ab 0 1 2
Time to immediate flow (min)c N/A 3 N/Ab N/A 3 5
Clot retrieved Yes Yes Yes Yes Yes Yes
No. passes with device 1 1 1 2 1 1
Final TIMI post-clot retrieval 3 3 3 3 3 3
Time to successful recanalization (min)d 7 6 9 35 4 8
Distal embolization No No No No No No
Vessel damage No No No No No No
Vasospasm Present Present Present Present Present Present
Thrombus No No No No No No
a

TIMI flow immediately postdeployment of the device into the clot.

b

Angiogram not obtained after deployment of device.

c

Time from insertion of microcatheter into the guide catheter to immediate flow restoration.

d

Time from insertion of microcatheter to retrieval of clot into the guide catheter.