. 2010 Feb;31(2):226–229. doi: 10.3174/ajnr.A1807

Table 1:

CTDIvol as a function of gantry angulationa

Tilt CTDIvol (mGy) Tilt CTDIvol (mGy)
30 44.8 0 51.6
25 46.9 −5 51.6
20 48.6 −10 51.1
15 50.1 −15 50.1
10 51.1 −20 48.6
5 51.6 −25 46.9
a

Scanning parameters were identical for each angle. Positive tilt values correspond to the caudal direction; negative tilt values correspond to the cranial direction.